CrossCHECK Locking Screw in the following sizes: 3.0 x 8 mm, REF CCP-L3008; 3.0 x 10 mm, REF CCP-L3010; 3.0 x...
Wright Medical Technology, Inc. began a recall of CrossCHECK Locking Screw in the following sizes: 3.0 x 8 mm, REF CCP-L3008; 3.0 x 10 mm, REF CCP-L3010; 3.0 x 12 mm, REF CCP-L3012; 3.0 x 14 mm, REF CCP-L3014... on . FDA classified it as Class II and gives the reason as "These screws have a potential of passing through the plate in the case when torque is applied in excessive amounts, which may result in intraoperative malfunction". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2541-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Wright Medical Technology Inc
- Product
- CrossCHECK Locking Screw in the following sizes: 3.0 x 8 mm, REF CCP-L3008; 3.0 x 10 mm, REF CCP-L3010; 3.0 x 12 mm, REF CCP-L3012; 3.0 x 14 mm, REF CCP-L3014; 3.0 x 16 mm, REF CCP-L3016; 3.0 x 18 mm, REF CCP-L3018; 3.0 x 20 mm, REF CCP-L3020; 3.0 x 22 mm, REF CCP-L3022; 3.0 x 24 mm, REF CCP-L3024; 3.0 x 26 mm, REF CCP-L3026; 3.0 x 28 mm, REF CCP-L3028; 3.0 x 30 mm, REF CCP-L3030; 3.0 x 32 mm, REF CCP-L3032; 3.0 x 34mm, REF CCP-L3034; 3.0 x 36 mm, REF CCP-L3036; 3.0 x 38 mm, REF CCP-L3038; 3.0 x 40 mm, REF CCP-L3040; 3.5 x 8 mm, REF CCP-L3508; 3.5 x 10 mm, REF CCP-L3510; 3.5 x 12 mm, REF CCP-L3512; 3.5 x 14mm, REF CCP-L3514; 3.5 x 16 mm, REF CCP-L3516; 3.5 x 18 mm, REF CCP-L3518; 3.5 x 20 mm, REF CCP-L3520; 3.5 x 22 mm, REF CCP-L3522; 3.5 x 24 mm, REF CCP-L3524; 3.5 x 26 mm, REF CCP-L3526; 3.5 x 28 mm, REF CCP-L3528; 3.5 x 30 mm, REF CCP-L3530; 3.5 x 35 mm, REF CCP-L3532; 3.5 x 34 mm, REF CCP-L3534; 3.5 x 36 mm, REF CCP-L3536; 3.5 x 38 mm, REF CCP-L3538; 3.5 x 40 mm, REF CCP-L3540 The CROSSCHECK Plating System is indicated for stabilization and fixation of fresh fractures, revision procedures, joint fusion and reconstruction of small bones in the hand, feet wrist and ankles, fingers and toes. The system may be used in both pediatric and adult patients. The device is intended for single use.
- Product code
- HRS Plate, Fixation, Bone
- Reason for recall
- These screws have a potential of passing through the plate in the case when torque is applied in excessive amounts, which may result in intraoperative malfunction.
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- Wright sent an Urgent Medical Device Recall letter dated July 20, 2015, to all affected customers. Customers were asked to complete the attached response form confirming receipt of the notice and the status of any products on hand, and return a copy to Wright by email ([email protected] or fax (901)867-7401. Also, if customers have any product in their inventory the product should be sent back to Arlington, TN (ALM). For questions regarding this recall call 901-290-5662.
- Quantity in commerce
- 107 units
- Distribution
- Nationwide Distribution
- Product codes and lots
- CrossCHECK Locking Screw with the following codes: REF CCP-L3008: Lot Number 35000; REF CCP-L3010: Lot Number 35000; REF CCP-L3012: Lot Number 35000-6, 35000; REF CCP-L3014: Lot Number 35000-8, 35000; REF CCP-L3016: Lot Number 35000; REF CCP-L3018: Lot Number 35000; REF CCP-L3020: Lot Number 35000; REF CCP-L3022: Lot Numbers: 35000-16-1, 35000; REF CCP-L3024: Lot Number 35000; REF CCP-L3026: Lot Number 35000; REF CCP-L3028: Lot Number 35000; REF CCP-L3030: Lot Number 35000; REF CCP-L3032: Lot Number 35000; REF CCP-L3034: Lot Number 35000; REF CCP-L3036: Lot Number 35000; REF CCP-L3038: Lot Number 35000; REF CCP-L3040: Lot Number 35000; REF CCP-L3508: Lot Number 35004; REF CCP-L3510: Lot Numbers: 35004-3, 35004; REF CCP-L3512: Lot Numbers: 35004-6, 35004; REF CCP-L3514: Lot Numbers: 35004-7, 35004; REF CCP-L3516: Lot Number 35004; REF CCP-L3518: Lot Numbers: 35004-12, 35004-12-1, 35004; REF CCP-L3520: Lot Numbers: 35004-13, 35004; REF CCP-L3522: Lot Numbers: 35004-16-1, 35004; REF CCP-L3524: Lot Numbers: 35004-18-1, 35004; REF CCP-L3526: Lot Number 35004; REF CCP-L3528: Lot Number 35004; REF CCP-L3530: Lot Number 35004; REF CCP-L3532: Lot Number 35004; REF CCP-L3534: Lot Number 35004; REF CCP-L3536: Lot Number 35004; REF CCP-L3538: Lot Number 35004; REF CCP-L3540: Lot Number 35004
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2541-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2542-2015 | Class II | Terminated | Voluntary: Firm initiated |