Z-2781-2015Class II
NebuTech Nebulizer, Model number: Series 8660, 8960, 8961, 8966, 8984. Nebulizer used to generate aerosols...
Salter Labs / initiated 2015-08-04 / Terminated
Salter Labs began a recall of NebuTech Nebulizer, Model number: Series 8660, 8960, 8961, 8966, 8984. Nebulizer used to generate aerosols delivered directly to the patient for breathing... on . FDA classified it as Class II and gives the reason as "The Inspiratory valve cap on the nebulizer reservoir tower may be loose and may detach before or during treatment". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-2781-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Salter Labs
- Product
- NebuTech Nebulizer, Model number: Series 8660, 8960, 8961, 8966, 8984. Nebulizer used to generate aerosols delivered directly to the patient for breathing. Designed to be used in either a hospital or homecare environment.
- Product code
- CAF Nebulizer (Direct Patient Interface)
- Reason for recall
- The Inspiratory valve cap on the nebulizer reservoir tower may be loose and may detach before or during treatment.
- FDA root cause
- Component change control
- Action
- Salter Labs sent a Medical Device Recall letter dated July 24 2015. The letter identified the affected product, problem and actions to be taken. The letter states that all affected product can be returned to Salter Labs for replacement at no charge with product specifically inspected for a tighter fit on the inspiratory valve cap. Customers were instructed to complete and return the Return Response Form as soon as possible. For any questions contact Salter Labs Customer Support at 800-421-0024 or the local Salter Labs representative.
- Quantity in commerce
- 8,975
- Distribution
- Worldwide Distribution - US Nationwide and in the countries of Greece, UK, Belgium, Australia, and Switerzland.
- Product codes and lots
- Lot Numbers 042015 through 070615A Series 8660, 8960, 8961, 8966, 8984. Nebulizers (part numbers 8660-7, 8960-7, 8960TG-7, 8967-7, 8982-7, 8984-7), Inspiratory valve cap (part number 600116).
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2781-2015 | Class II | Terminated | Voluntary: Firm initiated |