Z-2779-2015Class III

The Leica ASP200S/ASP300S Tissue Processor. Pathology and Hematology Departments. The Leica ASP200S/ASP300S...

Leica Microsystems, Inc. / initiated 2015-08-05 / Terminated / root cause: manufacturing process

Leica Microsystems, Inc. began a recall of The Leica ASP200S/ASP300S Tissue Processor. Pathology and Hematology Departments. The Leica ASP200S/ASP300S Tissue Processor is an automated system used to... on . FDA classified it as Class III and gives the reason as "The Leica ASP200S/ASP300S Tissue Processor have an incorrect internal wiring to the remote alarm plug, which can cause the remote alarm to not function correctly in the case of an instrument failure". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2779-2015
Classification
Class III
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Leica Microsystems, Inc.
Product
The Leica ASP200S/ASP300S Tissue Processor. Pathology and Hematology Departments. The Leica ASP200S/ASP300S Tissue Processor is an automated system used to process tissue specimens for examination through fixation, dehydration and infiltration.
Product code
IEO Processor, Tissue, Automated
Reason for recall
The Leica ASP200S/ASP300S Tissue Processor have an incorrect internal wiring to the remote alarm plug, which can cause the remote alarm to not function correctly in the case of an instrument failure.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Leica Biosystems sent an Urgent Medical Device Recall Notification letter dated August 4, 2015, to all affected customers to provide instructions on what the customer should do with the recalled product. These include 1) As an immediate countermeasure, the firm recommends that the customers do not use the remote alarm function on the Leica ASP200S/ASP300S in its current state. 2) The customers local Leica representative will contact them to arrange for a Leica service engineer to visit their facility to correct the internal wiring on their tissue processor and to check the remote alarm function. 3) The Medical Device Recall Notification needs to be passed on to all those who need to be aware within the custromers organisation or to any organisation where the potentially affected devices have been transferred. Please maintain awareness of this Medical Device Recall Notification and resulting action to ensure effectiveness of the corrective action. 4) The letter asks the customer use the enclosed Medical Device Recall Notification Response Form to confirm that they have received and understand the Medical Device Recall Notification. it states "Please confirm receipt of this notice by signing and dating the attached Acknowledgement Form and faxing it back to us at 1-847-236-3747, or scan and email it to [email protected]"
Quantity in commerce
901
Distribution
Worldwide Distribution - US including U.S. including AZ, CA, CT, DE, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, MN, MO, MT, NC, NJ, NM, NY, OH, OR, PA, TX, VA, WA, WI, WV and District of Columbia. Internationall to AE, AT, AU , BD , BE, BF, BG,BR, CA, CL, CN, CO, CR, DE, DK, EC, EG, ES, ET, FI, FR, GB, HU, ID, IN, IT, JM, JP, KR, LK, MA, MT, MY, NL, NZ, PA, PE, PH, PL, PT, RO, RU, SA, SE, SG, TH, TR, TW, VN and ZA.
Product codes and lots
Product Code: Leica ASP200S/ASP300S with Serial Numbers: 5682; 5886; 5264; 5267; 5356; 5259; 5956; 5946; 5794; 5385; 5929; 5301; 5401; 5310; 5530; 5909; 5804; 5272; 5724; 5796; 5807; 5446; 5435; 5670; 5947; 5348; 5434; 5388; 5366; 6000; 6131; 5407; 6040; 5799; 5815; 5737; 5706; 5381; 5416; 5673; 5875; 5962; 6041; 6042; 5266; 5948; 5693; 6021; 5610; 5798; 5926; 5387; 5349; 5273; 6115; 5612; 5965; 5955; 5260; 5725; 5871; 5873; 5874; 5562; 6046; 5386; 5977 ;5889; 5398; 6062; 5609; 6015; 5734; 5403; 5289; 5426; 6149; 5663; 6045

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2779-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2779-2015Class IIITerminatedVoluntary: Firm initiated