Z-2743-2015Class I

PROFEMUR PLUS CoCr MODULAR NECK, REF: PHAC1254, SIZE LONG, ORIENTATION 8 degree VAR/VAL. 12/14 SLT Taper...

Microport Orthopedics Inc. / initiated 2015-08-04 / Terminated / root cause: design

MicroPort Orthopedics, Inc. began a recall of PROFEMUR PLUS CoCr MODULAR NECK, REF: PHAC1254, SIZE LONG, ORIENTATION 8 degree VAR/VAL. 12/14 SLT Taper. Orthopedic hip prosthesis component on . FDA classified it as Class I and gives the reason as "Unexpected rate of postoperative fractures resulting in the need for revision surgery". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2743-2015
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Microport Orthopedics Inc.
Product
PROFEMUR PLUS CoCr MODULAR NECK, REF: PHAC1254, SIZE LONG, ORIENTATION 8 degree VAR/VAL. 12/14 SLT Taper. Orthopedic hip prosthesis component.
Product code
MBL Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Porous
Reason for recall
Unexpected rate of postoperative fractures resulting in the need for revision surgery.
Root cause tag
design
FDA root cause
Device Design
Action
Micro Port Orthopedics sent a Voluntary Device Product Recall letter dated August 7, 2015, via Fed Ex to all affected customers. The letter identified the product, the problem, and action to be taken by the customer. Customers were instructed to locate and return effected product immediately. Customers with questions were instructed to call 1-866-872-0211. Customers were also asked to return the enclosed form immediately even if they do not have any affected product to return.
Quantity in commerce
10,489 units
Distribution
Worldwide Distribution - US (nationwide) and Internationally to AT, BE, CA, CN, CZ, DE, ES, FI, FR, GB, IT, KR, LT, LV, NL, NO, PE, PL, PT, RU, and ST.
Product codes and lots
All lot codes

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2743-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2743-2015Class ITerminatedVoluntary: Firm initiated