Z-0408-2016Class II

Philips Healthcare Ingenuity CT Computed Tomography X-ray system

Philips North America LLC / initiated 2015-04-14 / Terminated / root cause: software

Philips Electronics North America Corporation began a recall of Philips Healthcare Ingenuity CT Computed Tomography X-ray system on . FDA classified it as Class II and gives the reason as "During scans with specific protocol steps, the software unexpectedly sets the Axial ( 2 axis ) scan length to a different length than the user set in the plan". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0408-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Philips Healthcare Ingenuity CT Computed Tomography X-ray system
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
During scans with specific protocol steps, the software unexpectedly sets the Axial ( 2 axis ) scan length to a different length than the user set in the plan.
Root cause tag
software
FDA root cause
Radiation Control for Health and Safety Act
Action
Philips Healthcare Planned Action: 1. Philips Healthcare will contact customers and initiate a software update to correct the defect. 2. Philips Healthcare engineers will install the software and perform testing to ensure the software update was effective. 3. The customer notification letter which includes a statement that Philips Healthcare will without charge, remedy the defect or bring the product into compliance. CDRH approves the CAP subject to the following conditions: 1. Notification of all dealers and purchasers is to be made within 15 working days of receipt of this letter in the manner specified in 21 CFR 1003.21 and 1003.22. This office and the Food and Drug Administration (FDA) district office coordinator noted below are to be included in the notification. 2. Philips Healthcare will implement this CAP by April 1, 2016. If you need any further information or support concerning this issue, please contact your local Philips representative or local Philips Healthcare office.
Quantity in commerce
105
Distribution
Worldwide Distribution - USA (nationwide) Distribution and to the countries of : Argentina, Australia, Austria, Bangladesh, Belgium, Bolivia, Brazil, Canada, China, Colombia, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Gibraltar, Great Britain, Greece, Hong Kong, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Latvia, Lebanon, Libya, Lithuania, Malaysia, Martinique, Mayanmar, Mexico, Mozambique, Netherlands, New Zealand, Norway, Panama, Philippines, Poland, Portugal, Republic of Korea, Reunion, Russian Federation, Saudi Arabia, Singapore, South Africa, Spain, State of Palestine, Sweden, Switzerland, Taiwan, Thailand, Tanzania, Tunisia, Turkey, United Arab Emirates, Uzbekstan and Vietnam.
Product codes and lots
728326

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0408-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0405-2016Class IITerminatedVoluntary: Firm initiated
Z-0406-2016Class IITerminatedVoluntary: Firm initiated
Z-0407-2016Class IITerminatedVoluntary: Firm initiated
Z-0408-2016Class IITerminatedVoluntary: Firm initiated