IBC Vascular Loop Catalog Number CH111, CH112, CH113, CH114, CH115, CH116, CH118
International Biophysics Corp. began a recall of IBC Vascular Loop Catalog Number CH111, CH112, CH113, CH114, CH115, CH116, CH118 Product Usage: The IBC Vascular Loop is a single-use, disposable loop used in... on . FDA classified it as Class II and gives the reason as "It was discovered that the heat seal placed onto sterile pouches may not meet strength acceptance criteria". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-2424-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- International Biophysics Corp.
- Product
- IBC Vascular Loop Catalog Number CH111, CH112, CH113, CH114, CH115, CH116, CH118 Product Usage: The IBC Vascular Loop is a single-use, disposable loop used in occluding, retracting, and identifying arteries, veins, tendons, and nerves in surgical procedures.
- Product code
- DWS Instruments, Surgical, Cardiovascular
- Reason for recall
- It was discovered that the heat seal placed onto sterile pouches may not meet strength acceptance criteria.
- Root cause tag
- packaging
- FDA root cause
- Packaging process control
- Action
- The recalling firm began notifying their affacted customers on 7/2/2015
- Quantity in commerce
- 771 units
- Distribution
- Worldwide - United States Nationwide and the countries of: Germany, Norway, Turkey, Chile, Pakistan, and Republic of South Africa
- Product codes and lots
- 031715-3418, 050815-3513, 052115-3540, 033015-3435, 052115-3542, 031615-3417, 050815-3514, 052115-3543, 033015-3436, 050815-3515, 060915-3578, 041715-3478, 033015-3437, 050815-3516, 052215-3545, 042015-3482, 052115-3541
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2422-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2423-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2424-2015 | Class II | Terminated | Voluntary: Firm initiated |