LX3 Floor Stand for Endure Luxite Envision Optical System; Endure LuxOR Q-Vue
Alcon Research, Ltd. began a recall of LX3 Floor Stand for Endure Luxite Envision Optical System; Endure LuxOR Q-Vue. Product Usage: The LX3 Microscope Floor Stand is an accessory to the LuxOR... on . FDA classified it as Class II and gives the reason as "Alcon is initiating a voluntary recall of all LX3 Microscope Floor Stands after receiving reports concerning loss of functionality due to issues with the cabling between the power, host and illumination modules...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2448-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Alcon Laboratories
- Product
- LX3 Floor Stand for Endure Luxite Envision Optical System; Endure LuxOR Q-Vue. Product Usage: The LX3 Microscope Floor Stand is an accessory to the LuxOR Opthalmic Microscope and the LuxOR Opthalmic Microscope with Q-VUE 3-D Assistant Visualization and provides a portable suspension system for the microscope that integrates an AC power module and microscope controls (Primary Control Panel and connectivity of foot controllers)
- Product code
- HRM Microscope, Operating & Accessories, Ac-Powered, Ophthalmic
- Reason for recall
- Alcon is initiating a voluntary recall of all LX3 Microscope Floor Stands after receiving reports concerning loss of functionality due to issues with the cabling between the power, host and illumination modules, potentially causing the system to display a system message, and/or resulting in illumination failure.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Alcon sent an Urgent Medical Device Correction letter dated August 11, 2015 to affected customers. The letter identified the affected product, problem and actions to be taken. For questions contact Alcon at 800-862-5266.
- Quantity in commerce
- 688 units
- Distribution
- Worldwide Distribution - US Nationwide and the countries Argentina, Australia, Austria, Belarus, Belgium, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Denmark, Dominican Republic, France, Germany, Greece, Guatemala, Hungary, India, Italy, Japan, Jordan, Korea, Lebanon, Malaysia, Mexico, Mongolia, New Zealand/Fiji, Oman, Panama, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Saudi Arabia, Singapore, Slovakia, South Africa, Spain, Switzerland, Taiwan, Thailand, Turkey, Ukraine and United Kingdom.
- Product codes and lots
- All units LX3 Microscope Floor Stands manufactured prior to July 2015 (Serial Numbers 1402548501X to 1502857606X)
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2448-2015 | Class II | Terminated | Voluntary: Firm initiated |