Z-0128-2016Class II
Villa Sistemi Rotograph EVO 3D X-ray
Acteon, Inc. / initiated 2015-08-05 / Terminated
Villa Radiology Systems LLC began a recall of Villa Sistemi Rotograph EVO 3D X-ray on . FDA classified it as Class II and gives the reason as "It has been discovered that due to lack of information accompanying documents the product fails to comply with 21CfR 1020.33(c),1020.33(d),1020.33(g),and 1020.330)". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0128-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Acteon, Inc.
- Product
- Villa Sistemi Rotograph EVO 3D X-ray
- Product code
- MUH System, X-Ray, Extraoral Source, Digital
- Reason for recall
- It has been discovered that due to lack of information accompanying documents the product fails to comply with 21CfR 1020.33(c),1020.33(d),1020.33(g),and 1020.330).
- FDA root cause
- Other
- Action
- Villa Radiology Systems Planned Action: 1. The current users manuals have been updated and addendums to the user manuals have been developed for currently installed Rotograph Evo 3D and Imax Touch 3D. The revisions and addendums contain all information required under 21 CFR 1020.33(c), 1020.33(d), 1020.33(g), and 1020.33(j). 2. New Rotograph Evo 3D and Imax Touch 3D systems will ship with the updated user manual. 3. Addendums will be provided to all end users CDRH approves the CAP subject to the following conditions: 1. Notification of all dealers and purchasers is to be made within 15 working days of receipt of this letter in the manner specified in 21 CFR 1003.21 and 1003.22.
- Quantity in commerce
- Rotograph EVO 3D - 8 installed
- Distribution
- US Distribution
- Product codes and lots
- p/n 9307251291, 9307351291, 9307651291
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0128-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0129-2016 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.