Z-0129-2016Class II

Owandy Radiology I -MAX Touch 3D X-ray

Acteon, Inc. / initiated 2015-08-05 / Terminated

Villa Radiology Systems LLC began a recall of Owandy Radiology I -MAX Touch 3D X-ray on . FDA classified it as Class II and gives the reason as "It has been discovered that due to lack of information accompanying documents the product fails to comply with 21CfR 1020.33(c),1020.33(d),1020.33(g),and 1020.330)". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0129-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Acteon, Inc.
Product
Owandy Radiology I -MAX Touch 3D X-ray
Product code
OAS X-Ray, Tomography, Computed, Dental
Reason for recall
It has been discovered that due to lack of information accompanying documents the product fails to comply with 21CfR 1020.33(c),1020.33(d),1020.33(g),and 1020.330).
FDA root cause
Other
Action
Villa Radiology Systems Planned Action: 1. The current users manuals have been updated and addendums to the user manuals have been developed for currently installed Rotograph Evo 3D and Imax Touch 3D. The revisions and addendums contain all information required under 21 CFR 1020.33(c), 1020.33(d), 1020.33(g), and 1020.33(j). 2. New Rotograph Evo 3D and Imax Touch 3D systems will ship with the updated user manual. 3. Addendums will be provided to all end users CDRH approves the CAP subject to the following conditions: 1. Notification of all dealers and purchasers is to be made within 15 working days of receipt of this letter in the manner specified in 21 CFR 1003.21 and 1003.22.
Quantity in commerce
I-MAX Touch 3D- 6 installed
Distribution
US Distribution
Product codes and lots
p/n 9306251191 and 9306651191

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0129-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0128-2016Class IITerminatedVoluntary: Firm initiated
Z-0129-2016Class IITerminatedVoluntary: Firm initiated