Centricity PACS-IW with Universal Viewer
GE Healthcare began a recall of Centricity PACS-IW with Universal Viewer Product Usage: The Centricity PACS-IW with Universal Viewer is a picture archiving and communications system, a device... on . FDA classified it as Class II and gives the reason as "Images from the Centricity PACS-IW with Universal Viewer and Centricity Universal Viewer may be missing when a system parameter MapRoute is set to a value greater than 1". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2469-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- Centricity PACS-IW with Universal Viewer Product Usage: The Centricity PACS-IW with Universal Viewer is a picture archiving and communications system, a device that provides one or more capabilities relating to the acceptance, transfer, display, storage, and digital processing of medical images. Its hardware components may include workstations, digitizers, communications devices, computers, video monitors, magnetic, optical disk, or other digital data storage devices, and hardcopy devices. The software components may provide functions for performing operations related to image manipulation, enhancement, compression or quantification. Used in Radiology.
- Product code
- LLZ System, Image Processing, Radiological
- Reason for recall
- Images from the Centricity PACS-IW with Universal Viewer and Centricity Universal Viewer may be missing when a system parameter MapRoute is set to a value greater than 1.
- Root cause tag
- software
- FDA root cause
- Software design (manufacturing process)
- Action
- The firm sent out an Urgent Medical Device Correction letter dated 8/11/2015 to all affected customers. Letter states that No User activity required. GE Healthcare has completed a remote inspection of your system and found that this system parameter is currently set to the default setting of 1. The firm will correct all affected systems by a providing software update. The letter also states that if the customer has any questions they should call GE Healthcare Service at 1-800-437-1171 or their local service representative.
- Quantity in commerce
- 82 systems
- Distribution
- Worldwide Distribution - U.S. nationwide in the states of: AK, CA, CO, DE, FL, HI, ID, NY, PA, SD, TN and TX; France, Germany, Greece, India, Israel, Malaysia, Mexico, Northern Ireland (UK), Panama, Poland, Russia, South Africa, Saudi Arabia, Switzerland, Turkey, UAE and Uruguay.
- Product codes and lots
- Model Number: Centricity PACS-IW with Universal Viewer Version 5.0 through 5.0 SP7 Codes: 2090255-001 UV 6.0 Web Client Media Kit. 05 Jun 15 to present
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2469-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2470-2015 | Class II | Terminated | Voluntary: Firm initiated |