Z-2773-2015Class III

Pointe Scientific Creatinine Reagent Catalog number C7539-150 in vitro diagnostic C7539-150 kit includes one...

Horiba Instruments Incorporated / initiated 2014-12-12 / Terminated / root cause: manufacturing process

Pointe Scientific, Inc. began a recall of Pointe Scientific Creatinine Reagent Catalog number C7539-150 in vitro diagnostic C7539-150 kit includes one R1 and one R2 bottle. R1: HDPE plastic bottle. 120... on . FDA classified it as Class III and gives the reason as "Creatinine kit C7539-150 contains Creatinine reagent R1 that is labeled as Creatinine reagent R2 (30ml) and reagent R2 vial is labeled as R1 (120ml). With manual procedure, no result will be reported as the control will...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2773-2015
Classification
Class III
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Horiba Instruments Incorporated
Product
Pointe Scientific Creatinine Reagent Catalog number C7539-150 in vitro diagnostic C7539-150 kit includes one R1 and one R2 bottle. R1: HDPE plastic bottle. 120 ml fill. R2: HDPE plastic bottle. 90 ml fill.
Product code
CGX Alkaline Picrate, Colorimetry, Creatinine
Reason for recall
Creatinine kit C7539-150 contains Creatinine reagent R1 that is labeled as Creatinine reagent R2 (30ml) and reagent R2 vial is labeled as R1 (120ml). With manual procedure, no result will be reported as the control will be out of specifications.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
On 12/12/2014, Field Correction Notice notifications were sent to the affected distributors via electronic mail. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Any questions concerning the corrective action may be directed to Pointe Scientifics Technical Service Department at 1-800-757-5313.
Quantity in commerce
93 kits
Distribution
Worldwide distribution. US nationwide, Singapore, Kenya, Bolivia, Mexico, Greece, Guyana, and Philippines.
Product codes and lots
C7539-150 Lot: 406201 Batch: 279 Expiration date: 2016-03

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2773-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2773-2015Class IIITerminatedVoluntary: Firm initiated