Pointe Scientific Creatinine Reagent Catalog number C7539-150 in vitro diagnostic C7539-150 kit includes one...
Pointe Scientific, Inc. began a recall of Pointe Scientific Creatinine Reagent Catalog number C7539-150 in vitro diagnostic C7539-150 kit includes one R1 and one R2 bottle. R1: HDPE plastic bottle. 120... on . FDA classified it as Class III and gives the reason as "Creatinine kit C7539-150 contains Creatinine reagent R1 that is labeled as Creatinine reagent R2 (30ml) and reagent R2 vial is labeled as R1 (120ml). With manual procedure, no result will be reported as the control will...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2773-2015
- Classification
- Class III
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Horiba Instruments Incorporated
- Product
- Pointe Scientific Creatinine Reagent Catalog number C7539-150 in vitro diagnostic C7539-150 kit includes one R1 and one R2 bottle. R1: HDPE plastic bottle. 120 ml fill. R2: HDPE plastic bottle. 90 ml fill.
- Product code
- CGX Alkaline Picrate, Colorimetry, Creatinine
- Reason for recall
- Creatinine kit C7539-150 contains Creatinine reagent R1 that is labeled as Creatinine reagent R2 (30ml) and reagent R2 vial is labeled as R1 (120ml). With manual procedure, no result will be reported as the control will be out of specifications.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- On 12/12/2014, Field Correction Notice notifications were sent to the affected distributors via electronic mail. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Any questions concerning the corrective action may be directed to Pointe Scientifics Technical Service Department at 1-800-757-5313.
- Quantity in commerce
- 93 kits
- Distribution
- Worldwide distribution. US nationwide, Singapore, Kenya, Bolivia, Mexico, Greece, Guyana, and Philippines.
- Product codes and lots
- C7539-150 Lot: 406201 Batch: 279 Expiration date: 2016-03
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2773-2015 | Class III | Terminated | Voluntary: Firm initiated |