Z-2719-2015Class II

Syngo.plaza; picture archiving and communications system

Siemens Medical Solutions USA, Inc. / initiated 2015-07-06 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of Syngo.plaza; picture archiving and communications system Product Usage: Syngo.plaza is a Picture Archiving and Communication System (PACS) intended to display... on . FDA classified it as Class II and gives the reason as "Potential issue leading to data loss and patient data mix-up". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2719-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Syngo.plaza; picture archiving and communications system Product Usage: Syngo.plaza is a Picture Archiving and Communication System (PACS) intended to display, process, read, report, communicate, distribute, store and archive digital medical images. It supports the physician in diagnosis and treatment planning. Syngo.plaza also supports storage and archiving of DICOM Structured reports. In a comprehensive imaging syngo.plaza integrates Hospital/Radiology Information Systems (HIS/RIS) to enable customer specific workflows. Syngo.plaza optionally uses a variety of advanced postprocessing applications.
Product code
LLZ System, Image Processing, Radiological
Reason for recall
Potential issue leading to data loss and patient data mix-up
Root cause tag
software
FDA root cause
Software design
Action
Siemens sent a Customer Information letter, dated July 6, 2015, to customers The letter identified the affected product, problem and action to be taken. The letter informed cusotmers about the Hotfix software update that was being released.
Quantity in commerce
37
Distribution
US Nationwide Distribution - in the states of CA, OR, FL, MO, NY, WA, MD, MA, NY, WI, KS, MN and PA.
Product codes and lots
Model Number of device  10592457 with serial numbers: Serial 100544 100499 100600 100571 100263 200308 200309 200310 200311 200312 200313 200314 200315 200316 200317 200318 200319 200320 200321 200322 200323 200324 100146 200277 100177 100138 100174 100491 100492 100596 100198 100181 200251 200252 100301 100302 100539

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2719-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2719-2015Class IITerminatedVoluntary: Firm initiated