KimVent* Wet Pak* Closed Suction System for Adults, 14 F, DT, T-Piece (Product Code 220569) A...
Halyard Health, Inc began a recall of KimVent* Wet Pak* Closed Suction System for Adults, 14 F, DT, T-Piece (Product Code 220569) A tracheobronchial suction catheter is a device that is attached to... on . FDA classified it as Class II and gives the reason as "The thumb valve of some KIMVENT Closed Suction Systems remained partially open after suctioning was completed, even though the valve appeared to be closed or in the "LOCKED" position". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2494-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Halyard Health
- Product
- KimVent* Wet Pak* Closed Suction System for Adults, 14 F, DT, T-Piece (Product Code 220569) A tracheobronchial suction catheter is a device that is attached to a port of the airway circuit for use as needed to aspirate liquids or semisolids from a patient's upper airway.
- Product code
- BSY Catheters, Suction, Tracheobronchial
- Reason for recall
- The thumb valve of some KIMVENT Closed Suction Systems remained partially open after suctioning was completed, even though the valve appeared to be closed or in the "LOCKED" position.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Halyard sent a Product Advisory Notice on July 29, 2015, via FEDEX with a second letter dated August 12, 2015, being issued via FEDEX. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to report any occurrence of a valve leak to Halyard Customer Service at 1-844-425-9273, option 1, option 3.
- Distribution
- Worldwide Distribution - US (nationwide and Puerto Rico and Internationally to Canada
- Product codes and lots
- Product Code 220569 - AB5145U14, AB5166U09, AB5180U11
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2492-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2493-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2494-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2495-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2496-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2497-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2498-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2499-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2500-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2501-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2502-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2503-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2504-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2505-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2506-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2507-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2508-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2509-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2510-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2511-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2512-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2513-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2514-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2515-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2516-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2517-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2518-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2519-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2520-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2521-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2522-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2523-2015 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.