Z-2487-2015Class II

Pointe Scientific G6PD Controls Kit configuration: 6 x 0.5 ml (2 vials each of Level 1, Level 2 and Level 3)...

Horiba Instruments Incorporated / initiated 2015-05-19 / Terminated / root cause: manufacturing process

Pointe Scientific, Inc. began a recall of Pointe Scientific G6PD Controls Kit configuration: 6 x 0.5 ml (2 vials each of Level 1, Level 2 and Level 3). Also sold as individual vials Bottle: Clear glass... on . FDA classified it as Class II and gives the reason as "A reduction in the reconstituted stability has been identified. Clinicians may notice a drop in recovery of the control after reconstitution, which could result in an out of specification result prior to the current...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2487-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Horiba Instruments Incorporated
Product
Pointe Scientific G6PD Controls Kit configuration: 6 x 0.5 ml (2 vials each of Level 1, Level 2 and Level 3). Also sold as individual vials Bottle: Clear glass vials Cap: Rubber seal with colored aluminum seals Glucose-6-phosphate dehydrogenase (G6PD) controls can be used to test for the quantitative and qualitative determination of G6PD in blood.
Product code
JBF Glucose-6-Phosphate Dehydrogenase (Erythrocytic), Screening
Reason for recall
A reduction in the reconstituted stability has been identified. Clinicians may notice a drop in recovery of the control after reconstitution, which could result in an out of specification result prior to the current reconstitution claim of 7 days. For this reason we suggest discontinuing use of the control set.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
The firm, POINTE SCIENTIFIC, issued a "TECHNICAL BULLETIN" on 5/19/2015, notifications were sent to the affected distributors via electronic mail. The recall notification included a description of the reason for the recall, affected product, and actions to be taken. The customers were instructed to: discontinue use of the control set and to incorporate this bulletin into your Quality Control files. The firm noted that this product has been placed on indefinite back-order and a replacement product is not currently available. For questions contact Pointe Scientific's Technical Service Department at 1-800-757-5313.
Quantity in commerce
2772 total vials (374 kits of 6 vials, 528 invidually sold vials)
Distribution
Worldwide distribution: US (nationwide) including states of: PA, CA, KY, FL, NJ, DE, MI, MN,and IL; and countries of: United Arab Emirates, Austria, Italy, Israel, Greece, and Australia.
Product codes and lots
Lot: 505801 Expiration date: 2015-11 catalog numbers: G7583-CTL 7-G7583-CTL-L1 7-G7583-CTL-L2 7-G7583-CTL-L3

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2487-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2487-2015Class IITerminatedVoluntary: Firm initiated