MEDTRONIC PERFUSION TUBING PACKS (with potentially affected Hemoconcentrators). Model # 7U65R2, 8C72R4...

Medtronic Perfusion Systems / initiated 2015-08-18 / Terminated / root cause: manufacturing process

Medtronic Perfusion Systems began a recall of MEDTRONIC PERFUSION TUBING PACKS (with potentially affected Hemoconcentrators). Model # 7U65R2, 8C72R4, 6N21R2, BB7M38R11, TL2Q68R17, HY8R66R3, HY2U43R9... on . FDA classified it as Class II and gives the reason as "Hemocor High Performance Hemoconcentrators demonstrate low ultrafiltration performance that is below product specification. These potentially affected units have been distributed as stand-alone devices and in specific...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2736-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic Perfusion Systems
Product
MEDTRONIC PERFUSION TUBING PACKS (with potentially affected Hemoconcentrators). Model # 7U65R2, 8C72R4, 6N21R2, BB7M38R11, TL2Q68R17, HY8R66R3, HY2U43R9, 8C72R4, BB5J59R19, 6B65R3, TL7M46R7, TL2Q68R17, 7U65R2, M074704B, M230002A, BB7M38Rll, M450013B, BB8N86R4, HY2U43R9, 8C72R4, M650123A, HY8R55R ,TL7M46R7, M971406C, M450013B, 7U65R2, M394415A, M450013B. Enables the perfusionist to manage the patient's hematocrit and fluid status.
Product code
DWE Tubing, Pump, Cardiopulmonary Bypass
Reason for recall
Hemocor High Performance Hemoconcentrators demonstrate low ultrafiltration performance that is below product specification. These potentially affected units have been distributed as stand-alone devices and in specific lots of Medtronic Perfusion Tubing packs.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Consignees were sent a Medtronic "Urgent Medical Device Recall" letter dated August 2015. The letter was addressed to "Risk Manager". The letter described the problem and the product involved in the recall. Informed consignees that no action is needed for product that has been used, but patients should continue to be managed in accordance with their standard patient management protocol. For unused product they recommend to immediately quarantine and return the affected product to Medtronic. Contact Customer Service at 800-854-3570 and reference RADAR #253277 to initiated a return and credit of product. Requested consignees to complete the Customer Confirmation Certificate and fax it to Medtronic at 651-367-0612 to the attention of Customer Focused Quality. If the product has been forwarded to another facility, consignees are advised to notify that facility. For questions contact Medtronic Lifeline Technical Services at 877-526-7890 or your Medtronic representative.
Quantity in commerce
Total 1594 (1306 US, 288 OUS)
Distribution
Worldwide Distribution-US (nationwide) including the states of AL, AR, CA, CO, FL, HI, ID, IN, KS, LA, MA, MS, MO, NY, NJ, OH, PA, TN, TX, UT, and WI, and the countries of Bolivia, Denmark, Dominican Republic, Italy, Latvia, Lebanon, Mexico, Panama, Singapore, Venezuela, and Zambia.
Product codes and lots
209050007 209087876 209107465 209135076 209162502 209195930 209241496 209241613 209270018 209270092 209297994 209298091 209298094 209310403 209326025 209373545 209388464 209396017 209401269 209401312 209415313 209431890 209464644 209512240 209556873 209621268 209644548 209653827

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2736-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2735-2015Class IITerminatedVoluntary: Firm initiated
Z-2736-2015Class IITerminatedVoluntary: Firm initiated