510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2780-2015Class II

ThruPort Knot Pusher. This device is sold individually (Model KP1) and as part of the ThruPort Valve...

Edwards Lifesciences LLC / initiated 2015-08-26 / Terminated / root cause: manufacturing process

Edwards Lifesciences, LLC began a recall of ThruPort Knot Pusher. This device is sold individually (Model KP1) and as part of the ThruPort Valve Placement Pack (Model VPP). packaged sterile and... on . FDA classified it as Class II and gives the reason as "The configuration of the slot at the tip of the knot pusher may inhibit intended ease of use. Fraying or splitting of the suture could occur, which could compromise the integrity of the suture knot". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2780-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Edwards Lifesciences LLC
Product
ThruPort Knot Pusher. This device is sold individually (Model KP1) and as part of the ThruPort Valve Placement Pack (Model VPP). packaged sterile and non-pyrogenic in a sealed, peel-type pouch. The Knot Pusher is a sterile disposable plastic rod used to advance extracorporeal suture knots during minimally invasive cardiac surgery.
Product code
DPT Probe, Blood-Flow, Extravascular
Reason for recall
The configuration of the slot at the tip of the knot pusher may inhibit intended ease of use. Fraying or splitting of the suture could occur, which could compromise the integrity of the suture knot.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Edwards sent an Urgent-Product Recall - Action Required letter dated August 26, 2015, to all affected customers. Letters were sent via Fed-Ex. Please check your inventory for the affected Lot Numbers listed above and return the enclosed confirmation form. If your inventory includes affected product, please contact Edwards Lifesciences to obtain an RGA number and return the affected product. We apologize for the inconvenience caused by this action and appreciate your attention to this matter. If you have questions that have not been answered by this letter, please call Edwards Customer Service at (800) 424-3278 from 6:00AM 4:30PM Pacific Time.
Quantity in commerce
759 units
Distribution
Worldwide Distribution - US (nationwide) Distribution and to the countries of : Canada, Japan and Europe.
Product codes and lots
Model: KP1 - Lot No: 59964698, 59967596, 59984375, 60016312, 60045801, 60045802, 60059982, 60066591, 60089865  Model VPP Lot No: 60022521, 60022522, 60064112

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2780-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2780-2015Class IITerminatedVoluntary: Firm initiated