Access Free T3 Reagent Kit, P/N A13422 The Access Free T3 assay is a paramagnetic particle, chemiluminescent...

Siemens Healthcare Diagnostics Inc. / initiated 2015-08-31 / Terminated

Beckman Coulter Inc. began a recall of Access Free T3 Reagent Kit, P/N A13422 The Access Free T3 assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of... on . FDA classified it as Class II and gives the reason as "Beckman Coulter, Inc. is recalling Access Free T3 Reagent Kit because they were filled incorrectly and contain insufficient quantity of reagents in one or more of the pack wells". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0110-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Access Free T3 Reagent Kit, P/N A13422 The Access Free T3 assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of free triiodothyronine levels in human serum and plasma using the Access Immunoassay Systems.
Product code
CDP Radioimmunoassay, Total Triiodothyronine
Reason for recall
Beckman Coulter, Inc. is recalling Access Free T3 Reagent Kit because they were filled incorrectly and contain insufficient quantity of reagents in one or more of the pack wells.
FDA root cause
Under Investigation by firm
Action
Beckman Coulter sent an " Urgent Medical Device Recall" letter dated August 31, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The letter informs the customers that some Access Free T3 reagent packs (p/n A13422) of lot 431433 were filled incorrectly and contain insufficient quantity of reagents in one or more of the pack wells. Customers are instructed to complete and return the enclosed response form within 10 days. Customers with any questions regarding the notice are instructed to contact Customer Support Center at http://ww.beckmancoulter.com; call (800) 854-3633 in US and Canada. Customers outside of US and Canada are instructed to contact their local Beckman Coulter representative.
Quantity in commerce
16,648 units total (1874 units in US)
Distribution
Worldwide Distribution - US (nationwide) and to the countries of : Australia, Brazil, Canada, China, Colombia, Costa Rica, Hong Kong, Israel, Japan, Kuwait, Macao, Malaysia, Mexico, New Zealand, Saudi Arabia, Singapore, Taiwan, Thailand and Viet Nam.
Product codes and lots
Lot: 431433

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0110-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0110-2016Class IITerminatedVoluntary: Firm initiated