Z-0019-2016Class II

SOMATOM Definition Edge; intended to produce cross-sectional images of the body by computer reconstruction of...

Siemens Medical Solutions USA, Inc. / initiated 2015-08-14 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of SOMATOM Definition Edge; intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same... on . FDA classified it as Class II and gives the reason as "software bug issues for SW-Version VA48A_SP0. The following safety issues were resolved: 1) Correction to improve visual warning and error indication son the gantry display. 2) Correction to improve acquisition data in...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0019-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
SOMATOM Definition Edge; intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes taken at different angles.
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
software bug issues for SW-Version VA48A_SP0. The following safety issues were resolved: 1) Correction to improve visual warning and error indication son the gantry display. 2) Correction to improve acquisition data in order to optimize image quality. 3) Correction to improve robustness and general system behavior in some exception handling procedures. 4) Correction to improve auto post processin
Root cause tag
software
FDA root cause
Software design
Action
A customer advisory notice, dated 8/14/15, was sent to direct accounts advising them of the issue to observe the safety notice and comply with the corresponding measure until the update has been fully completed.
Quantity in commerce
113 total
Distribution
Nationwide Distribution.
Product codes and lots
Model#' 10590000 with Serial numbers 83277 83281 83021 83004 83014 83276 83279

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0019-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0019-2016Class IITerminatedVoluntary: Firm initiated
Z-0020-2016Class IITerminatedVoluntary: Firm initiated
Z-0021-2016Class IITerminatedVoluntary: Firm initiated