Z-0228-2016Class II

MATResponder Tourniquet; Catalog number MATR: UPC 78353 00008; MATCombat Tourniquet, Catalog number MATC; UPC...

Pyng Medical Corp. / initiated 2015-08-31 / Terminated / root cause: manufacturing process

Pyng Medical Corporation began a recall of MATResponder Tourniquet; Catalog number MATR: UPC 78353 00008; MATCombat Tourniquet, Catalog number MATC; UPC 78353 00010; MATResponder Tourniquet and... on . FDA classified it as Class II and gives the reason as "Once tested, it is required that the tourniquet be returned to its Ready position then placed in its packaging. Each tourniquet was tested for functionality, but in some cases, the tourniquet was not returned to its...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0228-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Pyng Medical Corp.
Product
MATResponder Tourniquet; Catalog number MATR: UPC 78353 00008; MATCombat Tourniquet, Catalog number MATC; UPC 78353 00010; MATResponder Tourniquet and MATCombat Tourniquet are essentially the same product with different color schemes manual nonpneumatic tourniquet devices. The device provides for emergency blood flow occlusion to the limbs
Product code
GAX Tourniquet, Nonpneumatic
Reason for recall
Once tested, it is required that the tourniquet be returned to its Ready position then placed in its packaging. Each tourniquet was tested for functionality, but in some cases, the tourniquet was not returned to its Ready position prior to packaging.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
PYNG Medical sent Recall Notification letters dated August 28, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed: Please immediately cease use of the MATResponder and MATCombat tourniquet lots specified above. Inform all of your teams to cease use of or distribution of these lots. Please immediately place all MATResponder Tourniquets and MATCombat Tourniquets that you have in inventory into a quarantined location. For further questions please call (604) 303-7964. .
Quantity in commerce
6728
Distribution
Worldwide Distribution - US Distribution and to the countries of : Colombia, France, Canada, Italy, Germany, Finland, United Kingdom, Australia, The Netherlands, Japan, Hungary, Indonesia, Israel, Mexico, Poland and Spain.
Product codes and lots
MATR: lot numbers; SL115802 (Exp 07/2024), SL12136 (Exp 08/2024), SL121360 (Exp 12/2024), SL114711 (Exp 01/2024) MATC - lot numbers : SL12131 (Exp 07/2024), SL121312 (Exp 08/2024), SL113171 (Exp 10/2023).

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0228-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0228-2016Class IITerminatedVoluntary: Firm initiated