Specialist 2 Intramedullary (SP2 IM) Rod 400mm Instrument (PN 96-6120) The SP2 IM Rod 400mm Instrument (PN...

Depuy Orthopaedics Inc. / initiated 2015-09-08 / Terminated / root cause: design

DePuy Orthopaedics, Inc. began a recall of Specialist 2 Intramedullary (SP2 IM) Rod 400mm Instrument (PN 96-6120) The SP2 IM Rod 400mm Instrument (PN 96-6120) is used in both primary and revision P.F.C... on . FDA classified it as Class II and gives the reason as "DePuy Orthopaedics, Inc. is initiating a voluntary recall of select lots of the Specialist 2 Intramedullary Rod 400mm Instrument (PN 96-6120) due to complaints of the rod fracturing and leaving a portion of the rod in...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0111-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Depuy Orthopaedics Inc.
Product
Specialist 2 Intramedullary (SP2 IM) Rod 400mm Instrument (PN 96-6120) The SP2 IM Rod 400mm Instrument (PN 96-6120) is used in both primary and revision P.F.C. SIGMA knee procedures to align the femoral locating device and distal femoral cutting block. It is also used with the IM tibial resection. This rod is included with the P.F.C. SIGMA Specialist 2 (SP2 / SPII) Primary Knee Instruments, the P.F.C. SIGMA Specialist 2 (SP2 / SPII) Revision Knee Instruments, and P.F.C. SIGMA High Performance (HP) Primary Knee Instruments
Product code
LXH Orthopedic Manual Surgical Instrument
Reason for recall
DePuy Orthopaedics, Inc. is initiating a voluntary recall of select lots of the Specialist 2 Intramedullary Rod 400mm Instrument (PN 96-6120) due to complaints of the rod fracturing and leaving a portion of the rod in patients.
Root cause tag
design
FDA root cause
Device Design
Action
DePuy Synthes sent an Urgent Information Recall Notice to all affected customers. Instruments are not lot traceable in the system, so the company sent notices by e-mail on September 8, 2015, and US Distributors will hand deliver notices to all Medical Facilities that may have purchased the affected instrument. The US Distributor or US Sales Consultant will deliver the Medical Professional Notice and Reconciliation Form to the affected Medical Facilities. If assistance is needed, Sales Consultants will assist customers with inspection of all instruments, return of affected lots, receiving credit and returning the completed Medical Facility Reconciliation Forms to DePuy Orthopaedics, Inc. The affected instruments will be returned to the quarantine warehouse located at 700 Orthopaedic Drive, Warsaw, IN 46582. For questions regarding this recall call 574-371-4917. For questions regarding this recall call 574-267-8143.
Quantity in commerce
7488
Distribution
Worldwide Distribution - US including AL; AZ; CA; CO; CT; DC; FL; GA; HI; IA; IL; IN; KY; LA; MA; MD; ME; MI; MN; MO; MS; MT; NC; NE; NH; NJ; NV; NY; OH; OR; PA; RI; SC; TN; TX; UT; VA; VT; WA; WI; and WV and Internationally to ARGENTINA; AUSTRALIA; AUSTRIA; BELGIUM; BRAZIL; CANADA; CHILE; COLOMBIA; CZECH REPUBLIC; EGYPT; FRANCE; GERMANY; GREECE; INDIA; ISRAEL; ITALY; KOREA; Mexico; NEW ZEALAND; NORWAY; P R CHINA; PANAMA; PERU; POLAND; RUSSIA; SINGAPORE; SLOVENIA; SOUTH AFRICA; SWITZERLAND; UAE; UK; URUGUAY; VENEZUELA; and VIETNAM.
Product codes and lots
SP2 IM Rod 400mm Instruments (96-6120) manufactured with 455 stainless steel (SS) or 17-4 SS; CATALOG NO. : 96-6120; GTIN NO.: 10603295246893; Lots: Manufactured with 455 Stainless Steel, Labeled Lot #s-C98CS4; DE5P34; DE5RP4; D95AN4; EB5FV4; FA4G94; DE1FA4; DF4H44; DG9LK4; EB5GH4; EC9JY4; EF4DJ4; FD8MP4; FH8JA4; DG9L64; DJ5E34; DK3E34; EJ7AP4; DK3FE4; EJ7A34; ES2G64; ES2HA4; FH8JX4; ES2HY4; EX5L44; C3JHN4; EX5MS4; C3JHS4; E2SD44; C3JHF4; C4GA54; C4GBT4; C4GCC4; C52F14; C52F74; C52GL4; C52GV4; C67N14; C98BF4; TBACC; TBACZ; FJ4E74; TBCOJ; & Manufactured with 17-4 Stainless Steel- Etched Lot #s- H0797; H0100; H0199; H0200; H0298; H0301; H0396; H0399; H0401; H0495; H0499; H0530; H0596; H0598; H0600; H0601; H0696; H0699; H0796; H0797; H0800; H0899; H0900; H0997; H0998; H1095; H1097; H1195; H1198; H1297; H1298; H1299

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0111-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0111-2016Class IITerminatedVoluntary: Firm initiated