Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0122-2016Class III

Alere Cholestech LDX Multianalyte Controls, PN 88769 and 88773. Enables users to monitor the performance of...

Alere San Diego, Inc. / initiated 2015-08-27 / Terminated / root cause: labeling and UDI

Alere San Diego, Inc. began a recall of Alere Cholestech LDX Multianalyte Controls, PN 88769 and 88773. Enables users to monitor the performance of total cholesterol (TC), high density lipoprotein... on . FDA classified it as Class III and gives the reason as "Alere San Diego is recalling the Alere Cholestech LDX Multianalyte Control because the assigned control ranges for the Total Cholesterol (TC) and Triglyceride (TRG) analytes are incorrect". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0122-2016
Classification
Class III
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Alere San Diego, Inc.
Product
Alere Cholestech LDX Multianalyte Controls, PN 88769 and 88773. Enables users to monitor the performance of total cholesterol (TC), high density lipoprotein cholesterol (HDL), triglycerides (TRG) and glucose (GLU) test procedures.
Product code
JJY Multi-Analyte Controls, All Kinds (Assayed)
Reason for recall
Alere San Diego is recalling the Alere Cholestech LDX Multianalyte Control because the assigned control ranges for the Total Cholesterol (TC) and Triglyceride (TRG) analytes are incorrect.
Root cause tag
labeling and UDI
FDA root cause
Error in labeling
Action
Customer notification letters dated 8/25/15 and 9/3/15 were sent to customers to inform them that Alere San Diego is recalling the Alere Cholestech LDX Multianalyte Control because the assigned control ranges for the Total Cholesterol (TC) and Triglyceride (TRG) analytes are incorrect. Customers are informed that the Alere Cholestech LDX Total Cholesterol and Triglyceride assays continue to perform as expected. However, if customers' facility has used the control lot to test Total Cholesterol or Triglycerides as part of their Quality Control program, Alere San Diego recommends that they verify previous results or perform control testing using the replacement Expected Value Card or replacement control materials. The letters also provide customers with actions to be taken. Customers with questions are instructed to contact: Alere San Diego, Inc. 9975 Summers Ridge Road San Diego, CA 92121 U.S.A. Phone: 877-308-8289, FAX: 866-333-9839, E-mail: [email protected].
Quantity in commerce
190 kits
Distribution
Nationwide Distribution.
Product codes and lots
Lot C3091 and Lot C3091A

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0122-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0122-2016Class IIITerminatedVoluntary: Firm initiated