Z-0107-2016Class II

SOMATOM Force; computed tomography x-ray system. Intended to generate and process cross-sectional images of...

Siemens Medical Solutions USA, Inc. / initiated 2015-08-18 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of SOMATOM Force; computed tomography x-ray system. Intended to generate and process cross-sectional images of patients on . FDA classified it as Class II and gives the reason as "Software and firmware bugs". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0107-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
SOMATOM Force; computed tomography x-ray system. Intended to generate and process cross-sectional images of patients.
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
Software and firmware bugs
Root cause tag
software
FDA root cause
Software design
Action
Customers will be notified via the system interface and e-mail or U.S. postal mailing via a Customer Advisory Notice. If necessary the advisory letter may be hand delivered. The letter informed customers that the safety related issues identified were resolved by SW-Version VA50A_SP2. The update will be remotely pushed to affected systems. The letter also advised customers to notify anyone in their organization that should be aware of the information. In the event that the device had been sold, the notice should be forwarded to the new owner, and the firm should be informed of the new owner.
Quantity in commerce
21
Distribution
Nationwide Distribution
Product codes and lots
Model Number of device- 10742326  with serial numbers: 75514 75512 75510 75454 75437 75500 75475 75481 75439 75450 75482 75460 75493 75476 75458 75467 75487 75478 75513 75515, and  75501.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0107-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0107-2016Class IITerminatedVoluntary: Firm initiated