510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2811-2015Class II

Synthes Cranial Flap Tube Clamp and Crimping Device

Synthes (USA) Products LLC / initiated 2015-08-31 / Terminated / root cause: labeling and UDI

Synthes (USA) Products LLC began a recall of Synthes Cranial Flap Tube Clamp and Crimping Device Product Usage: The Synthes Cranial Flap Tube Clamp System is intended to reattach a cranial bone flap to... on . FDA classified it as Class II and gives the reason as "The Synthes Cranial Flap Tube Clamp and Crimping Device for Cranial Tube Clamp had been labeled MR Safe although they do not have the testing protocols currently required to designate them as MR Safe". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-2811-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Synthes (USA) Products LLC
Product
Synthes Cranial Flap Tube Clamp and Crimping Device Product Usage: The Synthes Cranial Flap Tube Clamp System is intended to reattach a cranial bone flap to the surrounding cranium after a craniotomy procedure.
Product code
GXN Plate, Cranioplasty, Preformed, Non-Alterable
Reason for recall
The Synthes Cranial Flap Tube Clamp and Crimping Device for Cranial Tube Clamp had been labeled MR Safe although they do not have the testing protocols currently required to designate them as MR Safe.
Root cause tag
labeling and UDI
FDA root cause
Labeling False and Misleading
Action
DePuy Synthes sent an URGENT:FIELD SAFETY NOTIFICATION / MEDICAL DEVICE LABELING CORRECTION letter, dated August 31, 2015 to affected customers. The letter identified affected product, problem and actions to be taken. Customers were instructed to: "Update and review the package insert (GP1345-D). " Discard outdated package insert noted in the table above. " Forward this Field Safety Notification to anyone in your facility that needs to be informed, especially personnel that conduct MR testing. " If the package insert has been forwarded to another facility, contact that facility. " Complete the attached Verification Section (page 3 of this notification). Please include your name, title, address, telephone number and signature in the spaces provided. Send a copy of the completed Verification Section by, Fax: 877-884-7821 or Scan/email: [email protected] " If the Verification Form is answered on behalf of more than one facility and/or individual, please clearly indicate the name and address of the facility and/or individual in page three (3) of this notification. " Maintain a copy of this notice. For questions call 610-719-5450.
Quantity in commerce
105882
Distribution
US Nationwide and internationally, including Canada.
Product codes and lots
Part Number 329.315 460.001 460.001.01S 460.002 460.002.01S 460.003 460.003.01S 460.008 460.008.01S 460.009 460.009.01S 460.010 460.010.01S 460.100 460.100.01S 460.107 460.107.01S with Lot Numbers 4146716, 4154144, 4206682, 4206683, 4213557, 4214824, 4216603, 4216604, 4240689, 4240690, 4241741, 4248462, 4249395, 4249396, 4264386, 4278302, 4278304, 4283035, 4283036, 4283037, 4283038, 4283039, 4283040, 4283041, 4285956, 4289103, 4295084, 4295085, 4298044, 4303093, 4326975, 4326979, 4326981, 4329355, 4331263, 4331265, 4331266, 4344659, 4344678, 4378973, 4482201, 4485317, 4488093, 4488094, 4490608, 4490609, 4496716, 4496915, 4496916, 4501971, 4504043, 4508068, 4513458, 4513459, 4513460, 4513461, 4516371, 4519406, 4519407, 4521408, 4525905, 4550062, 4550063, 4550064, 4550065, 4551164, 4551165, 4551166, 4561400, 4561401, 4572176, 4574481, 4580884, 4586704, 4605094, 4661414, 4669482, 4669483, 4675725, 4695343, 4705970, 4723148, 4753792, 4753794, 4777999, 4778000, 4906228, 4906229, 4914498, 4917681, 4934562, 4944809, 4956361, 4956363, 4983255, 4983256, 4983257, 5007806, 5041856, 5041857, 5041858, 5046857, 5046859, 5098159, 5113496, 5113497, 5134868, 5134869, 5143883, 5204105, 5230669, 5234395, 5234396, 5245233, 5245237, 5251710, 5292433, 5308690, 5308733, 5341509, 5370371, 5370375, 5438364, 5452452, 5465222, 5655435, 5667092, 5697537, 5726966, 5726967, 5726968, 5726969, 5726970, 5726971, 5801281, 5801284, 5801285, 5801287, 6036379, 6036381, 6168066, 6188888, 6188890, 6268389, 6268390, 6272370, 6322541, 6364587, 6553805, 6861415, 6983782, 7032568, 7505227, 7640627

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2811-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2811-2015Class IITerminatedVoluntary: Firm initiated