Z-0163-2016Class II

ARCHITECT i2000SR System The Abbott ARCHITECT System is intended for In Vitro diagnostic use only. The...

Abbott Medical / initiated 2015-09-09 / Terminated / root cause: manufacturing process

Abbott Laboratories, Inc began a recall of ARCHITECT i2000SR System The Abbott ARCHITECT System is intended for In Vitro diagnostic use only. The ARCHITECT system is designed to perform automated... on . FDA classified it as Class II and gives the reason as "Incorrect tubing". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0163-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
ARCHITECT i2000SR System The Abbott ARCHITECT System is intended for In Vitro diagnostic use only. The ARCHITECT system is designed to perform automated immunoassay tests, utilizing CMIA (chemiluminescent microparticle assay) detection technology.
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
Incorrect tubing.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Abbott Laboratories executed a Technical Service Bulletin on August 14, 2015.. The firm sent out a Product Correction notice on September 9, 2015 to all affected customers. The notice identified the product, the problem, and the action to be taken by the customer. Customers were instructed that an Abbott representative will make all necessary arrangements for replacing tubing on the impacted instruments. Customers were asked to retain a copy of the notice for their records. Customers with questions were instructed to contact Customer Service at 1-877-4ABBOTT. Customers outside the US should contact their local area Customer Service. For questions regarding this recall call 877-422-2688.
Quantity in commerce
65 units
Distribution
Worldwide Distribution - US (nationwide) and Internationally to France, Netherlands, Greece, Italy, Belgium, India, Austria, Germany, and Japan
Product codes and lots
ISR03967, ISR04737, ISR01461, ISR01998, ISR01964, ISR02446, ISR02751, ISR02781, ISR05266, ISR04520, ISR01460, ISR01442, ISR01988, ISR03034, ISR01194, ISR01583, ISR01955, ISR02437, ISR03144, ISR03352, ISR04369, ISR02360, ISR02361, ISR02842, ISR03495, ISR03756, ISR04244, ISR04420, ISR05155, ISR05240, ISR05242, ISR05433, ISR05671, ISR05883, ISR05946, ISR06050, ISR06054, ISR06837, ISR06864, ISR01696, ISR02073, ISR02264, ISR02269, ISR02999, ISR03668, ISR03901, ISR04053, ISR04165, ISR04388, ISR04586, ISR04619, ISR04770, ISR04773, ISR04775, ISR04801, ISR04996, ISR05202, ISR05662, ISR05732, ISR05751, ISR06406, ISR06592, ISR07108, ISR07240, ISR07741

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0163-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0163-2016Class IITerminatedVoluntary: Firm initiated