AltiVate Humeral Socket Shell Trial, Part Number 804-06-052 The glenoid baseplate is intended for cementless...
Encore Medical, Lp began a recall of AltiVate Humeral Socket Shell Trial, Part Number 804-06-052 The glenoid baseplate is intended for cementless application with addition of screws for fixation... on . FDA classified it as Class II and gives the reason as "Difficulty assembling the shell trial to the broach and the retaining ring disassembled from the screw". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0131-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Encore Medical, LP
- Product
- AltiVate Humeral Socket Shell Trial, Part Number 804-06-052 The glenoid baseplate is intended for cementless application with addition of screws for fixation. The humeral stem is intended for cemented or cementless use. This device may also be indicated in the salvage of previously failed surgical attempts. Shoulder reconstructive surgery should be reserved for patients who have failed non-operative treatments.
- Product code
- PHX Shoulder Prosthesis, Reverse Configuration
- Reason for recall
- Difficulty assembling the shell trial to the broach and the retaining ring disassembled from the screw.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- The firm, DJO Global, sent an "URGENT FIELD SAFETY NOTICE" letter dated 9/16/15 to affected consignees/customers. The letter described the product, problem, and actions to be taken. The consignees/customers were instructed quarantine the device, pass the notice on to all who need to be aware within your organization, contact customer service to place a replacement order and receive an RMA number; and to complete a response form and return the recalled device for a replacement. If you have any questions, please call (512) 834-6302 or by email at [email protected].
- Quantity in commerce
- 20 units
- Distribution
- US Distribution to states of: PA, FL, ID, and CA.
- Product codes and lots
- Lot: 180052L01
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0131-2016 | Class II | Terminated | Voluntary: Firm initiated |