Z-0112-2016Class II
Monaco Radiation Treatment Planning System. Used to make treatment plans for patients with prescriptions for...
Elekta Limited / initiated 2015-09-15 / Terminated / root cause: design
Elekta, Inc. began a recall of Monaco Radiation Treatment Planning System. Used to make treatment plans for patients with prescriptions for external beam radiation therapy on . FDA classified it as Class II and gives the reason as "Dose and MU are incorrect when CT images are viewed from the head, and, when using multiple prescriptions with forced densities". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0112-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Elekta Limited
- Product
- Monaco Radiation Treatment Planning System. Used to make treatment plans for patients with prescriptions for external beam radiation therapy.
- Product code
- MUJ System, Planning, Radiation Therapy Treatment
- Reason for recall
- Dose and MU are incorrect when CT images are viewed from the head, and, when using multiple prescriptions with forced densities.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- The firm issued an URGENT IMPORTANT FIELD SAFETY NOTICE on 09/15/2015. The letter identified the affected device, as well as the reason for the recall, and clinical impact. Customers were informed that both issues would be resolved in Monaco Patch Release 5.10.01. The attached acknowledgement form should be completed, signed, and returned to a local Elekta Office or representative within 30 days. Questions regarding the notice should be directed to a local Elekta office.
- Quantity in commerce
- 154 units
- Distribution
- Worldwide Distribution -- US, including the states of CA, IN, MI, MO, TX and WI; and, the countries of Australia, Canada, Germany, Greece, India, Netherlands, New Zealand, Turkey, and the UK.
- Product codes and lots
- Software Version 5.10.00
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0112-2016 | Class II | Terminated | Voluntary: Firm initiated |