Z-0112-2016Class II

Monaco Radiation Treatment Planning System. Used to make treatment plans for patients with prescriptions for...

Elekta Limited / initiated 2015-09-15 / Terminated / root cause: design

Elekta, Inc. began a recall of Monaco Radiation Treatment Planning System. Used to make treatment plans for patients with prescriptions for external beam radiation therapy on . FDA classified it as Class II and gives the reason as "Dose and MU are incorrect when CT images are viewed from the head, and, when using multiple prescriptions with forced densities". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0112-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Elekta Limited
Product
Monaco Radiation Treatment Planning System. Used to make treatment plans for patients with prescriptions for external beam radiation therapy.
Product code
MUJ System, Planning, Radiation Therapy Treatment
Reason for recall
Dose and MU are incorrect when CT images are viewed from the head, and, when using multiple prescriptions with forced densities.
Root cause tag
design
FDA root cause
Device Design
Action
The firm issued an URGENT IMPORTANT FIELD SAFETY NOTICE on 09/15/2015. The letter identified the affected device, as well as the reason for the recall, and clinical impact. Customers were informed that both issues would be resolved in Monaco Patch Release 5.10.01. The attached acknowledgement form should be completed, signed, and returned to a local Elekta Office or representative within 30 days. Questions regarding the notice should be directed to a local Elekta office.
Quantity in commerce
154 units
Distribution
Worldwide Distribution -- US, including the states of CA, IN, MI, MO, TX and WI; and, the countries of Australia, Canada, Germany, Greece, India, Netherlands, New Zealand, Turkey, and the UK.
Product codes and lots
Software Version 5.10.00

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0112-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0112-2016Class IITerminatedVoluntary: Firm initiated