Boston Scientific SQ-RX Model 1010 subcutaneous pulse generator, (S-ICD). Intended to provide defibrillation...

Boston Scientific Corporation / initiated 2015-09-17 / Terminated / root cause: manufacturing process

Boston Scientific Corporation began a recall of Boston Scientific SQ-RX Model 1010 subcutaneous pulse generator, (S-ICD). Intended to provide defibrillation therapy on . FDA classified it as Class II and gives the reason as "Devices shipped to US in dual-channel rather than US approved single-channel RF Telemetry communication mode. While this does not impact therapy delivery of device; this can impact likelihood of successfully...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0232-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Boston Scientific Corporation
Product
Boston Scientific SQ-RX Model 1010 subcutaneous pulse generator, (S-ICD). Intended to provide defibrillation therapy.
Product code
LWS Implantable Cardioverter Defibrillator (Non-Crt)
Reason for recall
Devices shipped to US in dual-channel rather than US approved single-channel RF Telemetry communication mode. While this does not impact therapy delivery of device; this can impact likelihood of successfully establishing telemetry during first scan of device when using a single-channel programmer. Telemetry can be established with additional scans; once established no further difficulty is expecte
Root cause tag
manufacturing process
FDA root cause
Process design
Action
One consignee was contacted via a Boston Scientific "Medical Device Retrieval" letter dated September 17, 2015. The letter was addressed to Hospital Administrator. The letter described the problem and the product involved in the recall. Informed consignee that their local Boston Scientific sales representative will be retrieving the product form their inventory. For questions they can contact their local sales representative or Boston Scientific Technical service at 1-800-227-3422. A second Boston Scientific "Important Medical Device Information" letter dated October 2015 was hand delivered to physicians starting on 10/15/2015. The letter described the problem and provided information on Clinical Considerations, Affected Population, Recommendations and Further Information. For questions they can contact their Boston scientific representative or the Americas Technical service at 1-800-227-3422.
Quantity in commerce
3
Distribution
Distributed in the US to MA and PR.
Product codes and lots
A019995 A020132 A020240

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0232-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0232-2016Class IITerminatedVoluntary: Firm initiated