Synthes Screw Inserter; for use with Intramedullary Rod, orthopedic device. Used to secure the head element...
Synthes (USA) Products LLC began a recall of Synthes Screw Inserter; for use with Intramedullary Rod, orthopedic device. Used to secure the head element to the head element inserter to prevent... on . FDA classified it as Class II and gives the reason as "Certain lots of the Screw Inserter may be etched with incorrect graphics. The orientation of the etched image of the oblique angle on the Screw Inserter may not correspond to the tip of the angle of the instrument". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0126-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Synthes (USA) Products LLC
- Product
- Synthes Screw Inserter; for use with Intramedullary Rod, orthopedic device. Used to secure the head element to the head element inserter to prevent disengagement during head element insertion.
- Product code
- HSB Rod, Fixation, Intramedullary And Accessories
- Reason for recall
- Certain lots of the Screw Inserter may be etched with incorrect graphics. The orientation of the etched image of the oblique angle on the Screw Inserter may not correspond to the tip of the angle of the instrument.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- DePuy Synthes sent an urgent notice of medical device recall letter dated September 21, 2015, to all affected customers to notify them about the issue, the potential risks, and to provide instructions for return of the completed response form and affected devices. Customers with questions were instructed to call 610-719-5450.
- Quantity in commerce
- 793
- Distribution
- US Nationwide Distribution
- Product codes and lots
- Part Number 03.037.025 Lot Numbers 8911887; 8911975; 9166356; 9285766; 9286671; 9319408; 9375245; 9375251; 9388359; 9388360; 9393071; 9393074; 9393075; 9393076; 9409795; 9409796; 9422892; 9422895; 9498455; 9310899; 9310900; 9319407
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0126-2016 | Class II | Terminated | Voluntary: Firm initiated |