Z-0261-2016Class II
Stryker Orthopaedics Tapers LFIT V40 Vitallium femoral heads. hip prosthesis component
Howmedica Osteonics Corp / initiated 2015-09-08 / Terminated / root cause: manufacturing process
Stryker Howmedica Osteonics Corp. began a recall of Stryker Orthopaedics Tapers LFIT V40 Vitallium femoral heads. hip prosthesis component on . FDA classified it as Class II and gives the reason as "Complaints that the femoral head could not be assembled with its corresponding V40 stem trunnion at the time of surgery". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0261-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Howmedica Osteonics Corp
- Product
- Stryker Orthopaedics Tapers LFIT V40 Vitallium femoral heads. hip prosthesis component.
- Product code
- LPH Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
- Reason for recall
- Complaints that the femoral head could not be assembled with its corresponding V40 stem trunnion at the time of surgery.
- Root cause tag
- manufacturing process
- FDA root cause
- Process design
- Action
- Stryker sent an Urgent Medical Device Recall Notifiications dated September 9 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers are asked to complete the attached form and return within 5 days by using email [email protected] or FAX 1-888-965-5803. For further questions please call ( 201) 831-5000.
- Quantity in commerce
- 201
- Distribution
- Nationwide Distribution
- Product codes and lots
- Catalog Numbers 6260-5-028; 6260-5-032; 6260-5-132; 6260-5-232, 232; 6260-5-328; 6260-5-332; 6260-5-428; 6260-5-432; 6260-9-0928; 6260-9-032, 6260-9-132, 6260-9-232, 6260-9-328, 6260-9-332, 6260-9-428, 6280-0-128, 6280-0-228, 6280-0-232, 6280-0-332 exp- dates july 2019 to august 2019
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0261-2016 | Class II | Terminated | Voluntary: Firm initiated |