Eon 300 system includes the following product reference numbers and names: Reference no. 79345 Glucose (GPSL)...
ELITech Clinical Systems SAS began a recall of Eon 300 system includes the following product reference numbers and names: Reference no. 79345 Glucose (GPSL) Reference no. 79450 Uric acid (AUVD) Reference... on . FDA classified it as Class II and gives the reason as "ELITech clinical Systems reagents based on Trinder reaction used on the Eon 300 instrument system are recalled due to potential drug interference. No complaints were received". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0215-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- VitalScientific
- Product
- Eon 300 system includes the following product reference numbers and names: Reference no. 79345 Glucose (GPSL) Reference no. 79450 Uric acid (AUVD) Reference no. 79435 Triglycerides (TGML) Reference no. 79295 Cholesterol (CHVD) Reference no. 79301 Cholesterol HDL (HDLL) Reference no. 79310 Cholesterol, LDL (LDLL)
- Product code
- CGA Glucose Oxidase, Glucose
- Reason for recall
- ELITech clinical Systems reagents based on Trinder reaction used on the Eon 300 instrument system are recalled due to potential drug interference. No complaints were received.
- FDA root cause
- Under Investigation by firm
- Action
- The firm sent recall letters or Technical Bulletins to US consignees on 9/22/15 via registered mail. Technical Bulletin #1022 (for EON 100 users) Technical Bulletin #2008 (for EON 300 users) Technical Bulletin #5064 (for ENVOY users) Technical Bulletin #3009 (for Selectra users) ELITechGroup is not requesting return of the affected product. The method sheets and package inserts will be updated with the new interference claims. When available, the revised labeling will be provided to the customer for customer's use until the new IFU is included in the product packaging. Consignees are requested to return the Confirmation of Receipt form back to the firm by FAX to 401-642-9001 or send email to [email protected] or send via US Post: Vital Diagnostics, 27 Wellington Rd., Lincoln, Rl. If you have any questions, please contact Vital Diagnostics Technical Support at 1-800-345-2822 or 855-ELITECH
- Quantity in commerce
- 612 total
- Distribution
- US Nationwide Distribution
- Product codes and lots
- all lots
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0214-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0215-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0216-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0217-2016 | Class II | Terminated | Voluntary: Firm initiated |