Envoy 500 system includes the following product reference numbers and names: 55345 ENVOY 500 Glucose 55450...
ELITech Clinical Systems SAS began a recall of Envoy 500 system includes the following product reference numbers and names: 55345 ENVOY 500 Glucose 55450 ENVOY 500 URIC ACID REAGENT KIT 55435 ENVOY 500... on . FDA classified it as Class II and gives the reason as "ELITech clinical Systems reagents based on Trinder reaction used on the Envoy 500 instrument system are recalled due to potential drug interference. No complaints were received". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0217-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- VitalScientific
- Product
- Envoy 500 system includes the following product reference numbers and names: 55345 ENVOY 500 Glucose 55450 ENVOY 500 URIC ACID REAGENT KIT 55435 ENVOY 500 TRIGLYCERIDES REAGENT KIT, 55297 ENVOY 500 CHOLESTEROL REAGENT KIT, 55335 ENVOY 500 CREATININE REAGENT KIT 55301 ENVOY 500 HDL CHOLESTEROL REAGENT 55310 ENVOY 500 LDL CHOLESTEROL REAGENT KIT
- Product code
- CGA Glucose Oxidase, Glucose
- Reason for recall
- ELITech clinical Systems reagents based on Trinder reaction used on the Envoy 500 instrument system are recalled due to potential drug interference. No complaints were received.
- FDA root cause
- Under Investigation by firm
- Action
- The firm sent recall letters or Technical Bulletins to US consignees on 9/22/15 via registered mail. Technical Bulletin #1022 (for EON 100 users) Technical Bulletin #2008 (for EON 300 users) Technical Bulletin #5064 (for ENVOY users) Technical Bulletin #3009 (for Selectra users) ELITechGroup is not requesting return of the affected product. The method sheets and package inserts will be updated with the new interference claims. When available, the revised labeling will be provided to the customer for customer's use until the new IFU is included in the product packaging. Consignees are requested to return the Confirmation of Receipt form back to the firm by FAX to 401-642-9001 or send email to [email protected] or send via US Post: Vital Diagnostics, 27 Wellington Rd., Lincoln, Rl. If you have any questions, please contact Vital Diagnostics Technical Support at 1-800-345-2822 or 855-ELITECH
- Quantity in commerce
- 13455 systems
- Distribution
- US Nationwide Distribution
- Product codes and lots
- all lots
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0214-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0215-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0216-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0217-2016 | Class II | Terminated | Voluntary: Firm initiated |