Z-1066-2016Class II

Philips X-Ray Systems, Allura Xper with R8.2.16

Philips North America LLC / initiated 2015-10-01 / Terminated / root cause: software

Philips Electronics North America Corporation began a recall of Philips X-Ray Systems, Allura Xper with R8.2.16 Product Usage: The Allura Xper FD1O and Allura Xper FD1O/10 is intended for use in cardiovascular and vascular... on . FDA classified it as Class II and gives the reason as "Upon initiating Fluoroscopy the user may encounter a user message Fluoro failed". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1066-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Philips X-Ray Systems, Allura Xper with R8.2.16 Product Usage: The Allura Xper FD1O and Allura Xper FD1O/10 is intended for use in cardiovascular and vascular X-ray imaging applications, including diagnostic, interventional procedures (such as PTCA, stent placement and atherectomies), pacemaker implantations and Electrophysiology (EP). The Allura Xper FD2O, Allura Xper FD2O/10 and Allura Xper FD2O/20 is intended for: Dedicated vascular and neurovascular imaging applications, including diagnostic and interventional procedures.
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
Upon initiating Fluoroscopy the user may encounter a user message Fluoro failed.
Root cause tag
software
FDA root cause
Software design
Action
Philips sent a Field Safety Notice to all affected customers on October 1, 2015. The notice informs customers of the issue and the actions planned by Philips. Philips Field Service Engineers are to install Software R8.2.16.1 which addresses the issue. For further information or support concerning the issue, consignees should contact their local Philips Representative or call the Technical Support Line at 800-722-9377.
Quantity in commerce
196 devices
Distribution
US Nationwide in the states of : AK, AL, AZ, CA, FL, GA, HI, IL, IN, KS, KY, LA, MA, MD, MI, MO, MS, NC, NJ, NY, OH, OR, PA, TN, TX, UT, WI, WV and the country of Canada.
Product codes and lots
Product Codes: 722026, 722027, 722028, 722029, 722033, 722034, 722035, 722036, 722038, 722039, 722058, 722059 systems with release R8.2.16

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1066-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1066-2016Class IITerminatedVoluntary: Firm initiated