510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0221-2016Class II

SmartSite Extension Set, Model No. 20029E. Used to administer fluid and medications through a needle or...

CareFusion 303, Inc / initiated 2015-10-09 / Terminated / root cause: manufacturing process

CareFusion 303, Inc. began a recall of SmartSite Extension Set, Model No. 20029E. Used to administer fluid and medications through a needle or catheter inserted into the patient's artery or vein on . FDA classified it as Class II and gives the reason as "CareFusion is recalling the SmartSite Extension Set due to disconnection and leakage issues". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0221-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
CareFusion 303, Inc
Product
SmartSite Extension Set, Model No. 20029E. Used to administer fluid and medications through a needle or catheter inserted into the patient's artery or vein.
Product code
FPA Set, Administration, Intravascular
Reason for recall
CareFusion is recalling the SmartSite Extension Set due to disconnection and leakage issues.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
An urgent medical device recall notification was sent to customers on 10/9/15 who purchased the SmartSite Extension Set. The letter informs the customers that CareFusion has identified potential risks with model code 20029E, lot numbers 1503541, 15015769 and 15045950 due to disconnection and leakage issues. The letter also provides the customers with the potential risk involved and the actions to be taken. Customers are instructed to complete and return to CareFusion the recall response card. Customers with recall related questions are instructed to contact CareFusion Support Center at (888) 562-6018. Customers with adverse event reports are instructed to contact Customer Advocacy at (888) 812-3266 or email [email protected]. Customers with technical questions regarding technical questions are instructed to contact Technical Support at (888) 812-3229.
Quantity in commerce
21,000 units
Distribution
Distributed US (nationwide) and in Canada.
Product codes and lots
Lot No. 15015769, 15035412, 15045950.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0221-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0221-2016Class IITerminatedVoluntary: Firm initiated