Cyberonics Vagus Nerve Stimulator (VNS) Therapy AspireSR Generator, Model 106
Cyberonics, Inc began a recall of Cyberonics Vagus Nerve Stimulator (VNS) Therapy AspireSR Generator, Model 106 on . FDA classified it as Class II and gives the reason as "Certain Model 106 Pulse Generators demonstrate delays in sensing during use of the 'Verify Heartbeat Detection' feature and exhibit potential for decreased battery longevity". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0280-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- LivaNova USA, Inc.
- Product
- Cyberonics Vagus Nerve Stimulator (VNS) Therapy AspireSR Generator, Model 106
- Product code
- LYJ Stimulator, Autonomic Nerve, Implanted For Epilepsy
- Reason for recall
- Certain Model 106 Pulse Generators demonstrate delays in sensing during use of the 'Verify Heartbeat Detection' feature and exhibit potential for decreased battery longevity.
- Root cause tag
- manufacturing process
- FDA root cause
- Manufacturing material removal
- Action
- Cyberonics sent written recall notifications beginning October 14, 2015, to all affected customers. The letter identified the product the problem and the action to be taken by the customer. The notice also directs customers to contact Cyberonics to schedule additional on-site troubleshooting. The on-site troubleshooting proposed by Cyberonics will utilize a feature included within the Model 250 Version 9.0 Programming Software as an alternate method of verifying heart rate and determining that the appropriate heart beat sensitivity setting is programmed. The Version 9.0 software displays an instantaneous heart rate when the user selects the 'Refresh' feature. This heart rate is measured directly by the generator and displayed on the programmer. For further questions, please call (866) 882-8804.
- Quantity in commerce
- 27 units
- Distribution
- US Distribution to the states of : OR, NY, GA, PA, CA, NC, KS, FL, KY, MI, TX, IL, OH, IN, UT, CO and WV.
- Product codes and lots
- Serial Numbers: 37719, 37899, 38123, 38414, 38428, 39385, 39531, 39607, 39801, 39821, 40037, 40048, 40249, 40251, 40271, 40354, 40428, 40488, 40526, 40603, 40751, 41000, 41109, 41152, 41805, 42022, 42780
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0280-2016 | Class II | Terminated | Voluntary: Firm initiated |