Z-0280-2016Class II

Cyberonics Vagus Nerve Stimulator (VNS) Therapy AspireSR Generator, Model 106

LivaNova USA, Inc. / initiated 2015-10-07 / Terminated / root cause: manufacturing process

Cyberonics, Inc began a recall of Cyberonics Vagus Nerve Stimulator (VNS) Therapy AspireSR Generator, Model 106 on . FDA classified it as Class II and gives the reason as "Certain Model 106 Pulse Generators demonstrate delays in sensing during use of the 'Verify Heartbeat Detection' feature and exhibit potential for decreased battery longevity". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0280-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
LivaNova USA, Inc.
Product
Cyberonics Vagus Nerve Stimulator (VNS) Therapy AspireSR Generator, Model 106
Product code
LYJ Stimulator, Autonomic Nerve, Implanted For Epilepsy
Reason for recall
Certain Model 106 Pulse Generators demonstrate delays in sensing during use of the 'Verify Heartbeat Detection' feature and exhibit potential for decreased battery longevity.
Root cause tag
manufacturing process
FDA root cause
Manufacturing material removal
Action
Cyberonics sent written recall notifications beginning October 14, 2015, to all affected customers. The letter identified the product the problem and the action to be taken by the customer. The notice also directs customers to contact Cyberonics to schedule additional on-site troubleshooting. The on-site troubleshooting proposed by Cyberonics will utilize a feature included within the Model 250 Version 9.0 Programming Software as an alternate method of verifying heart rate and determining that the appropriate heart beat sensitivity setting is programmed. The Version 9.0 software displays an instantaneous heart rate when the user selects the 'Refresh' feature. This heart rate is measured directly by the generator and displayed on the programmer. For further questions, please call (866) 882-8804.
Quantity in commerce
27 units
Distribution
US Distribution to the states of : OR, NY, GA, PA, CA, NC, KS, FL, KY, MI, TX, IL, OH, IN, UT, CO and WV.
Product codes and lots
Serial Numbers: 37719, 37899, 38123, 38414, 38428, 39385, 39531, 39607, 39801, 39821, 40037, 40048, 40249, 40251, 40271, 40354, 40428, 40488, 40526, 40603, 40751, 41000, 41109, 41152, 41805, 42022, 42780

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0280-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0280-2016Class IITerminatedVoluntary: Firm initiated