Z-0798-2016Class I

Shuttle Select Slip-Cath Catheter. Angiographic catheter. For use in angiographic procedures as a conduit for...

Cook Incorporated / initiated 2015-10-07 / Terminated / root cause: design

Cook Inc. began a recall of Shuttle Select Slip-Cath Catheter. Angiographic catheter. For use in angiographic procedures as a conduit for the delivery of contrast media and can help... on . FDA classified it as Class I and gives the reason as "Reports of catheter tip splits and or separation. Loss of device function, medical intervention to retrieve separated segment, or complications resulting from separated tip occluding blood flow to end organs may result...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0798-2016
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Cook Incorporated
Product
Shuttle Select Slip-Cath Catheter. Angiographic catheter. For use in angiographic procedures as a conduit for the delivery of contrast media and can help physicians diagnose occlusion or stenosis. The catheters are designed for percutaneous introduction into the vascular system over an appropriately sized wire guide.
Product code
DQO Catheter, Intravascular, Diagnostic
Reason for recall
Reports of catheter tip splits and or separation. Loss of device function, medical intervention to retrieve separated segment, or complications resulting from separated tip occluding blood flow to end organs may result. Recall expanded 10/07/2015 to include additional catheters.
Root cause tag
design
FDA root cause
Component design/selection
Action
On 7/2/2015 Cook Incorporated sent, URGENT: MEDICAL DEVICE RECALL notifications dated July 2, 2015, to affected customers with instructions for returning the affected product. All customers were notified via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Should you have any questions or concerns, please do not hesitate to contact Cook Medical Customer Relations at (800) 457-4500 or (812) 339-2235. Firm issued press release 8/3/2015.On 10/09/2015 the firm expanded the recall and reissued press release via www.businesswire.com to include additional catheter products and lots as specified in the release.
Quantity in commerce
9519
Distribution
Worldwide Distribution - US (nationwide including Puerto Rico and US Virgin Islands), Algeria, Antigua, Armenia, Austria, Australia, Bahamas, Belgium, Bangladesh, Bulgaria, Bahrain, Brazil, Brunei Darussalam, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Dominican Republic, Egypt, El Salvador, Finland, France, French Polynesia, Georgia, Germany, Ghana, Great Britain, Greece Guadeloupe, Hong Kong, Hungary, Iceland, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jersey, Jordan, Kuwait, Kazakhstan, Republic of Korea, Kuwait, Lebanon, Lithuania, Luxemburg, Latvia, Libya, Macau, Malaysia, Malta, Martinique, Mexico, Mongolia, Morocco, Netherlands, Netherlands Antilles, New Caledonia, New Zealand, Nicaragua, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Sri Lanka, Sweden, Switzerland, Spain, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, and Vietnam.
Product codes and lots
All lot numbers of the following Product numbers:  SCBR4.5-35-135-P-NS-ANG-SHTL, SCBR4.5-35-135-P-NS-ANG-SHTL-JP, SCBR4.5-35-150-P-NS-0-SHTL, SCBR4.5-35-150-P-NS-ANG-SHTL, SCBR4.5-35-75-P-NS-ANGSHTL

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0798-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0798-2016Class ITerminatedVoluntary: Firm initiated