Z-0340-2016Class II

ADVIA Chemistry Hemoglobin A1c_3, A1c_3M Reagent kits. Used on the ADVIA 1200, 1650, 1800, 2400, and XPT...

Siemens Healthcare Diagnostics Inc. / initiated 2015-09-24 / Terminated

Siemens Healthcare Diagnostics, Inc. began a recall of ADVIA Chemistry Hemoglobin A1c_3, A1c_3M Reagent kits. Used on the ADVIA 1200, 1650, 1800, 2400, and XPT Chemistry Systems. For in vitro diagnostic use in the... on . FDA classified it as Class II and gives the reason as "Reagent in these lots may demonstrate an increased occurrence of high %HbA1c bias, a positive bias averaging 0.6% HbA1c units, ranging from -0.1% to 1.1% HbA1c units. The maximum bias was observed at higher %HbA1c...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0340-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
ADVIA Chemistry Hemoglobin A1c_3, A1c_3M Reagent kits. Used on the ADVIA 1200, 1650, 1800, 2400, and XPT Chemistry Systems. For in vitro diagnostic use in the quantitative determination of Hemoglobin A1c, a diabetes marker, in whole blood on the ADVIA Chemistry systems.
Product code
LCP Assay, Glycosylated Hemoglobin
Reason for recall
Reagent in these lots may demonstrate an increased occurrence of high %HbA1c bias, a positive bias averaging 0.6% HbA1c units, ranging from -0.1% to 1.1% HbA1c units. The maximum bias was observed at higher %HbA1c concentrations. QC samples may exhibit a similar bias.
FDA root cause
Under Investigation by firm
Action
Siemens has issued an Urgent Medical Device Recall in the US and an Urgent Field Safety Notice Outside the US on September 24, 2015, instructing customers to discontinue use, discard reagent and request no charge replacement kits for these products.
Quantity in commerce
3031
Distribution
Distributed in the states of AZ, NY, CA, IN, LA, MN, OH, MI, NM, and GA, and the countries of Mexico and Canada.
Product codes and lots
10379673 (160 tests kit), 10485591 (800 tests kit) Lot 230 (SMN 10485591 and 10379673) and lot 231(SMN 10485591)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0340-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0340-2016Class IITerminatedVoluntary: Firm initiated