Z-0181-2016Class II
Monaco - Radiation Treatment Planning used to make treatment plans for patients with prescriptions for...
Elekta Limited / initiated 2015-10-16 / Terminated
Elekta, Inc. began a recall of Monaco - Radiation Treatment Planning used to make treatment plans for patients with prescriptions for external beam radiation therapy on . FDA classified it as Class II and gives the reason as "Unintended update of Dose and MU and Incorrect Assignment of Bolus". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0181-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Elekta Limited
- Product
- Monaco - Radiation Treatment Planning used to make treatment plans for patients with prescriptions for external beam radiation therapy.
- Product code
- MUJ System, Planning, Radiation Therapy Treatment
- Reason for recall
- Unintended update of Dose and MU and Incorrect Assignment of Bolus.
- FDA root cause
- Other
- Action
- Elekta sent Important Field Safety Notice 382-01-MON-003 to all customers on 10/16/2015. The notice informs users of the specific product and version numbers affected by both issues. The affected devices for both issues are those running Monaco 5.10.01. Customers are instructed to complete and return acknowledgement form. Solution/fix has been released in Monaco patch 5.10.02 and all affected users will be made aware of its availability by the end of October 2015.
- Quantity in commerce
- 154
- Distribution
- CA, IN, MI,MO, TX, WI, Australia, Canada, Germany,m Greece, India, Netherlands, New Zealand, Turkey and United Kingdom
- Product codes and lots
- Software Version 5.10.01