Z-0297-2016Class II

Bulk Loader Module for cobas p 512/612 Pre-analytical sample handling that includes de-capping, aliquoting...

Roche Diagnostics GmbH / initiated 2015-09-22 / Terminated / root cause: design

Roche Diagnostics Operations, Inc. began a recall of Bulk Loader Module for cobas p 512/612 Pre-analytical sample handling that includes de-capping, aliquoting and sorting of samples for analysis. In vitro... on . FDA classified it as Class II and gives the reason as "Possible for sample tubes in the loader to be opened and contaminate the system as well as adjacent samples during processing". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0297-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Roche Diagnostics GmbH
Product
Bulk Loader Module for cobas p 512/612 Pre-analytical sample handling that includes de-capping, aliquoting and sorting of samples for analysis. In vitro diagnostic machine component.
Product code
JQP Calculator/Data Processing Module, For Clinical Use
Reason for recall
Possible for sample tubes in the loader to be opened and contaminate the system as well as adjacent samples during processing.
Root cause tag
design
FDA root cause
Device Design
Action
Roche Diagnostics Operations, Inc. sent an Urgent Medical Device Correction letter to affected consignees. The letter identified the product, the problem, and the action to be taken by the consignee. Consignees were asked to retain the Urgent Medical Device Correction letter until a Roche Field Service Representative has installed the BLM modification kit. Consignees were also asked to complete the attached form and fax to 1-877-906-8982, ext 2015. Consignees with questions were instructed to call 1-800-428-2336. For questions regarding this recall call 317-521-3911.
Quantity in commerce
1
Distribution
Distributed to TX only.
Product codes and lots
During transport from the Bulk Loader Module (BLM), catalog number 07135645001, to the cobas p 512/p 612 pre-analytical instruments, sample tubes may inadvertently open and spill, causing contamination of the system.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0297-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0297-2016Class IITerminatedVoluntary: Firm initiated