Bulk Loader Module for cobas p 512/612 Pre-analytical sample handling that includes de-capping, aliquoting...
Roche Diagnostics Operations, Inc. began a recall of Bulk Loader Module for cobas p 512/612 Pre-analytical sample handling that includes de-capping, aliquoting and sorting of samples for analysis. In vitro... on . FDA classified it as Class II and gives the reason as "Possible for sample tubes in the loader to be opened and contaminate the system as well as adjacent samples during processing". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0297-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Roche Diagnostics GmbH
- Product
- Bulk Loader Module for cobas p 512/612 Pre-analytical sample handling that includes de-capping, aliquoting and sorting of samples for analysis. In vitro diagnostic machine component.
- Product code
- JQP Calculator/Data Processing Module, For Clinical Use
- Reason for recall
- Possible for sample tubes in the loader to be opened and contaminate the system as well as adjacent samples during processing.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Roche Diagnostics Operations, Inc. sent an Urgent Medical Device Correction letter to affected consignees. The letter identified the product, the problem, and the action to be taken by the consignee. Consignees were asked to retain the Urgent Medical Device Correction letter until a Roche Field Service Representative has installed the BLM modification kit. Consignees were also asked to complete the attached form and fax to 1-877-906-8982, ext 2015. Consignees with questions were instructed to call 1-800-428-2336. For questions regarding this recall call 317-521-3911.
- Quantity in commerce
- 1
- Distribution
- Distributed to TX only.
- Product codes and lots
- During transport from the Bulk Loader Module (BLM), catalog number 07135645001, to the cobas p 512/p 612 pre-analytical instruments, sample tubes may inadvertently open and spill, causing contamination of the system.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0297-2016 | Class II | Terminated | Voluntary: Firm initiated |