Z-0365-2016Class II

Patient Support Material Number: 9890 010 87431, used with:EasyDiagnost ; Model: 706050

Philips North America LLC / initiated 2014-08-19 / Terminated / root cause: design

Philips Medical Systems, Inc. began a recall of Patient Support Material Number: 9890 010 87431, used with:EasyDiagnost ; Model: 706050 on . FDA classified it as Class II and gives the reason as "The hook does not securely hold the footplate in vertical position. Therefore the footplate can fall or swing down and may hurt the operator or patient". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0365-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Patient Support Material Number: 9890 010 87431, used with:EasyDiagnost ; Model: 706050
Product code
KPR System, X-Ray, Stationary
Reason for recall
The hook does not securely hold the footplate in vertical position. Therefore the footplate can fall or swing down and may hurt the operator or patient.
Root cause tag
design
FDA root cause
Device Design
Action
The Field Safety Notice FCO70600086 will be mailed on August 19, 2014. Philips Medical Systems will send out certified letters and will have the receipts returned for tracking purposes as well as have a Field Service Engineer visit each site to provide the fix when the Field Change Order is released.
Quantity in commerce
260 units, not all units have the patient support accessory
Distribution
Distributed in the states of MN, KY, MA, VA, TX, NC, HI, FL, AZ, CT, and WY and the countries of Argentina, Australia, Austria, Canada, Denmark, Germany, Netherlands, Russia, Saudi Arabia, Singapore, Spain, and Switzerland.
Product codes and lots
All patient support for stitching are affected.