Z-0344-2016Class II

Two-stage and three-stage I.V. poles are shipped as an optional accessory to a bed or stretcher. The I.V...

Stryker Medical / initiated 2015-10-13 / Terminated / root cause: manufacturing process

Stryker Medical Division of Stryker Corporation began a recall of Two-stage and three-stage I.V. poles are shipped as an optional accessory to a bed or stretcher. The I.V. poles are stationary, height adjustable stands... on . FDA classified it as Class II and gives the reason as "It was identified by a customer complaint that some I.V. poles were retracting/lowering from an extended position to the low height unexpectedly. An investigation was initiated which identified that the I.V. Latch...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0344-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Stryker Medical
Product
Two-stage and three-stage I.V. poles are shipped as an optional accessory to a bed or stretcher. The I.V. poles are stationary, height adjustable stands intended to hold infusion liquids, infusion accessories, and/or other medical devices. They may be used on the following parent devices: Model 2030 (EPIC II), model 3005 S3 beds and stretcher. models 1105, 1105E, 1105X, 1105XE, 1115, 1115E, 1115X, 1115XE, 1125E, 1125X and 1125XE.
Product code
FOX Stand, Infusion
Reason for recall
It was identified by a customer complaint that some I.V. poles were retracting/lowering from an extended position to the low height unexpectedly. An investigation was initiated which identified that the I.V. Latch Housing was nonconforming. potential hazards: " Suspended I.V. pole retracts unexpectedly " Suspended I.V. pole retracts unexpectedly with pump/bags attached
Root cause tag
manufacturing process
FDA root cause
Process control
Action
On 10/13/2015, URGENT MEDICAL DEVICE NOTIFICATION letters were sent to the affected customers via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. If you have any urgent questions or concerns, please, contact Stryker Customer Service at (1-800-327-0770). Our normal business hours are Monday-Friday 8 a.m-5 p.m. (ET).
Quantity in commerce
19,876
Distribution
Worldwide Distribution-US (nationwide) including DC and Guam, and the countries of New Zealand Canada Chile Hong Kong Colombia Argentina Brazil France Germany Italy Netherlands Spain Switzerland United Kingdom India Japan South Korea Mexico South Africa Poland Singapore.
Product codes and lots
Stryker 2-Stage and 3-Stage I.V. Poles distributed on beds, stretchers, and in service kits. This product field action is limited to I.V. Poles distributed between 01/07/2014 - 06/24/2014.  The following I.V. Pole option numbers come affixed to beds and stretchers:  Manufacturer Model Number and Product Description 2035-112-010; Head End Left, 2 Stage I.V. Pole 2035-113-011; Head End Right, 2 Stage I.V. Pole 0785-035-101; 26", 2 Stage IV Pole Assembly 0785-035-200; 26", 3 Stage IV Pole Assembly 0785-035-300; 30", 3 Stage IV Pole Assembly 0785-035-401; 30", 2 Stage IV Pole Assembly   The following I.V. Pole option numbers ship as service kits.  Manufacturer Service Kit Number and Product Description 1105-035-368; 30" IV POLE 2 STAGE HE PR 1105-035-367; 30" IV POLE 2 STAGE HE PL 1105-035-343; 30" IV POLE 2 STAGE FE PR 1105-035-341; 30" IV POLE, 2 STAGE HE, PR 1105-035-340; 30" IV POLE, 2 STAGE FE, PL 1105-035-338; 30" IV POLE, 2 STAGE FE, PL 1040-035-338; 2-STAGE IV POLEHEAD LT30" ZM 0785-035-340; 2-STAGE IV POLE,HEAD LT, 30" 0785-035-341; 2-STAGE IV POLEHEAD RT 30" 0785-035-343; 2-STAGE IV POLE,FOOT RT, 30" 1105-035-344; 30" IV POLE, 3 STAGE FE, PL 1105-035-342; 30" IV POLE, 3 STAGE HE, PL 1105-035-339; 30" IV POLE, 3 STAGE FE, PR 1105-035-369; 30" IV POLE, 3 STAGE HE, PL 1105-035-366; 30" IV POLE, 3 STAGE HE, PR 0785-035-339; 3-STAGE IV POLE,FOOT RT, 30" 0785-035-342; 3-STAGE IV POLE,HEAD LT, 30" 1105-035-644; 26" IV POLE, 3 STAGE FL 1105-035-642; 26" IV POLE, 3 STAGE HL 1105-035-639; 26" IV POLE 3 STAGE FR 1105-035-637; 26" IV POLE, 3 STAGE HR 0785-035-637; 3-STAGE IV POLE,HEAD RT, 26" 0785-035-639; 3-STAGE IV POLE,FOOT RT, 26" 0785-035-642; 3-STAGE IV POLE,HEAD LT, 26" 1105-035-643; 26" IV POLE, 2 STAGE FR 1105-035-641; 26" IV POLE, 2 STAGE HR 1105-035-640; 26" IV POLE, 2 STAGE FL 1105-035-638; 26" IV POLE, 2 STAGE HL 0785-035-638; 2-STAGE IV POLE,HEAD LT, 26" 0785-035-640; 2-STAGE IV POLE,FOOT LT, 26" 0785-035-641; 2-STAGE IV POLE,HEAD RT, 26" 0785-035-643; 2-STAGE IV POLE,FOOT RT, 26" 2035-112-000; IV POLE - 2 STAGE H/E 2035-113-000; DUAL H/E 2 STAGE IV POLE OPT

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0344-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0344-2016Class IITerminatedVoluntary: Firm initiated