Z-0225-2016Class II
The Centricity PACS-IW with Universal Viewer is a device that displays medical images (including mammograms)...
GE Healthcare / initiated 2015-09-28 / Terminated / root cause: software
GE Healthcare began a recall of The Centricity PACS-IW with Universal Viewer is a device that displays medical images (including mammograms) and data from various imaging sources. Images and... on . FDA classified it as Class II and gives the reason as "Inaccurate distance measurements with magnified projection X-ray images". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0225-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- The Centricity PACS-IW with Universal Viewer is a device that displays medical images (including mammograms) and data from various imaging sources. Images and data can be viewed, communicated, processed and displayed within the system or across computer networks at distributed locations. Typical users of this system are trained professionals, including but not limited to radiologists, physicians, nurses, medical technicians, and assistants.
- Product code
- LLZ System, Image Processing, Radiological
- Reason for recall
- Inaccurate distance measurements with magnified projection X-ray images.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- An URGENT MEDICAL DEVICE CORRECTION letter was sent to all medical users on 9/28/2015 regarding Centricity PACS-IW with Universal Viewer and Centricity Universal Viewer- Inaccurate distance measurements with magnified projection X-ray images. GE Healthcare explains that for all projection X-rays, GE Healthcare recommends that the user perform the necessary measurements on the standard contact images, and not rely on measurements made on the magnified views unless the measurements have been independently verified for each X-Ray machine installed or manually calibrated using a radio-opaque fiducial of a known size. Attention should be given to prior cases acquired before the application of the safety instructions given in this communication as they may be affected by the safety issue. GE Healthcare will correct all affected systems by providing a software update at no cost to the customer. GE Healthcare advises customers if they have any questions or concerns regarding the notification, to please contact GE Healthcare Service at 1-800-437-1171 or their local Service Representative.
- Quantity in commerce
- Centricity PACS-IW with Universal Viewer Versions 5.0 SP2 and higher 1052 units
- Distribution
- Worldwide Distribution-US (nationwide) including the states of AZ, CA, CO, DE, FL, GA, IA, IL, IN, LA, MA, MI, MN, MO, NJ, NY, OR, PA, SC, SD, TN, TX, VA, WA, WI and WV, and the countries of Australia, Austria, Bangladesh, Belgium, Brazil, Canada, Chile, China, Ecuador, France, Germany, Ghana, Hong Kong, India, Israel, Italy, Japan, Kenya, Korea, Malta, Malaysia, Namibia, Netherlands, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Turkey, Taiwan, United Arab Emirates, United Kingdom and Venezuela.
- Product codes and lots
- 2067131-001DVD UNIVERSAL VIEWER 5.0 - WEB CLIENT
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0225-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0226-2016 | Class II | Terminated | Voluntary: Firm initiated |