Z-0226-2016Class II

The Centricity Universal Viewer Version is a device that displays medical images (including mammograms) and...

GE Healthcare / initiated 2015-09-28 / Terminated / root cause: software

GE Healthcare began a recall of The Centricity Universal Viewer Version is a device that displays medical images (including mammograms) and data from various imaging sources. Images and data... on . FDA classified it as Class II and gives the reason as "Inaccurate distance measurements with magnified projection X-ray images". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0226-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
The Centricity Universal Viewer Version is a device that displays medical images (including mammograms) and data from various imaging sources. Images and data can be viewed, communicated, processed and displayed within the system or across computer networks at distributed locations. Typical users of this system are trained professionals, including but not limited to radiologists, physicians, nurses, medical technicians, and assistants.
Product code
LLZ System, Image Processing, Radiological
Reason for recall
Inaccurate distance measurements with magnified projection X-ray images.
Root cause tag
software
FDA root cause
Software design
Action
An URGENT MEDICAL DEVICE CORRECTION letter was sent to all medical users on 9/28/2015 regarding Centricity PACS-IW with Universal Viewer and Centricity Universal Viewer- Inaccurate distance measurements with magnified projection X-ray images. GE Healthcare explains that for all projection X-rays, GE Healthcare recommends that the user perform the necessary measurements on the standard contact images, and not rely on measurements made on the magnified views unless the measurements have been independently verified for each X-Ray machine installed or manually calibrated using a radio-opaque fiducial of a known size. Attention should be given to prior cases acquired before the application of the safety instructions given in this communication as they may be affected by the safety issue. GE Healthcare will correct all affected systems by providing a software update at no cost to the customer. GE Healthcare advises customers if they have any questions or concerns regarding the notification, to please contact GE Healthcare Service at 1-800-437-1171 or their local Service Representative.
Quantity in commerce
Centricity Universal Viewer versions 6.0 and higher 5 units
Distribution
Worldwide Distribution-US (nationwide) including the states of AZ, CA, CO, DE, FL, GA, IA, IL, IN, LA, MA, MI, MN, MO, NJ, NY, OR, PA, SC, SD, TN, TX, VA, WA, WI and WV, and the countries of Australia, Austria, Bangladesh, Belgium, Brazil, Canada, Chile, China, Ecuador, France, Germany, Ghana, Hong Kong, India, Israel, Italy, Japan, Kenya, Korea, Malta, Malaysia, Namibia, Netherlands, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Turkey, Taiwan, United Arab Emirates, United Kingdom and Venezuela.
Product codes and lots
2088034-001 DVD UNIVERSAL VIEWER 6.0 - WEB CLIENT

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0226-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0225-2016Class IITerminatedVoluntary: Firm initiated
Z-0226-2016Class IITerminatedVoluntary: Firm initiated