Z-0264-2016Class II

CARESCAPE VC150 Vital Signs Monitor Monitor vital signs in humans

Innokas Medical Oy / initiated 2015-10-22 / Terminated / root cause: software

Innokas Medical Oy began a recall of CARESCAPE VC150 Vital Signs Monitor Monitor vital signs in humans on . FDA classified it as Class II and gives the reason as "Monitor may shut down unintentionally without restarting". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0264-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Innokas Medical Oy
Product
CARESCAPE VC150 Vital Signs Monitor Monitor vital signs in humans
Product code
MWI Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms)
Reason for recall
Monitor may shut down unintentionally without restarting.
Root cause tag
software
FDA root cause
Software design
Action
The firm, Innokas Medical, sent a "Urgent Field Safety Notice(s)" dated October 22, 2015 and October 23, 2015 by e-mail to their consignees. The notice informed the distributor of the device, about the deficiency, and correction of the device. The notice included the instructions regarding what to do with the product and information that the devices will be updated. The consignees were instructed to complete and return the Medical Device FSCA/Recall Return Response Acknowledgement and Receipt Form via fax to +358 8 562 3151, mail to: Innokas Medical Oy, Vihikari 10, FI-90440-Kempele, Finland or email to: [email protected]. If you have any questions, contact the QA/RA Director at +358 44 562 3131.
Quantity in commerce
424 units (38 in US)
Distribution
Worldwide Distribution: US (nationwide) and Europe.
Product codes and lots
Catalog numbers: 2067980-001, 2067980-002, 2067980-003, 2067980-004, 2067980-005, 2067980-006, 2067980-007, 2067980-008, 2067980-009, 2067980-010, 2067980-011, 2067980-012, 2067980-013, 2067980-014, 2067980-015, 2067980-016  Serial Numbers: In USA: SK514470004YA,SK514470005YA,SK515050001YA,SK515050002YA,SK515050003YA,SK515050004YA, SK515050005YA,SK515050006YA,SK515050007YA,SK515050008YA,SK515050009YA,SK515050010YA, SK515050011YA,SK515050012YA,SK515050013YA,SK515050014YA,SK515050015YA,SK515050026YA, SK515050027YA,SK515050028YA,SK515050029YA,SK515050030YA,SK515050031YA,SK515050032YA, SK515050033YA,SK515050034YA,SK515050035YA,SK515050036YA,SK515050037YA,SK515050038YA, SK515050039YA,SK515050040YA,SK515050041YA,SK515050042YA,SK515050043YA,SK515050044YA, SK515050045YA,SK515100051YA  In Europe: SK514150001YA,SK514150004YA,SK514150005YA,SK514150007YA,SK514150008YA,SK514150009YA, SK514150010YA,SK514150011YA,SK514150019YA,SK514150020YA,SK514150023YA,SK514150026YA, SK514150027YA,SK514150028YA,SK514150029YA,SK514150030YA,SK514150031YA,SK514300002YA, SK514300003YA,SK514300004YA,SK514300005YA,SK514300006YA,SK514300007YA,SK514300008YA, SK514300009YA,SK514300010YA,SK514300011YA,SK514300014YA,SK514300015YA,SK514380001YA, SK514380002YA,SK514380003YA,SK514380004YA,SK514380005YA,SK514380006YA,SK514380007YA, SK514380008YA,SK514380009YA,SK514380010YA,SK514380011YA,SK514380012YA,SK514390001YA, SK514390005YA,SK514390006YA,SK514390007YA,SK514390008YA,SK514390009YA,SK514390010YA, SK514390011YA,SK514390012YA,SK514390013YA,SK514390014YA,SK514390016YA,SK514390017YA, SK514390018YA,SK514390019YA,SK514390020YA,SK514390021YA,SK514390023YA,SK514390024YA, SK514390025YA,SK514390026YA,SK514390027YA,SK514390028YA,SK514390029YA,SK514390030YA, SK514390031YA,SK514390032YA,SK514450001YA,SK514450002YA,SK514450003YA,SK514450004YA, SK514450005YA,SK514450006YA,SK514450008YA,SK514450011YA,SK514450018YA,SK514450021YA, SK514450022YA,SK514460003YA,SK514460004YA,SK514460005YA,SK514460006YA,SK514460007YA, SK514460008YA,SK514460009YA,SK514460010YA,SK514460011YA,SK514460012YA,SK514460013YA, SK514460014YA,SK514460015YA,SK514460016YA,SK514470006YA,SK514470007YA,SK514470008YA, SK514470009YA,SK514470010YA,SK514470011YA,SK514470012YA,SK514470013YA,SK514470014YA, SK514470015YA,SK514470016YA,SK514470018YA,SK514470020YA,SK514470025YA,SK514470026YA, SK514470027YA,SK514470028YA,SK514470029YA,SK514470030YA,SK514470031YA,SK514470032YA, SK514470033YA,SK514470034YA,SK514470035YA,SK514470036YA,SK514470037YA,SK514470038YA, SK514470039YA,SK514470040YA,SK514480003YA,SK514480004YA,SK514480005YA,SK514480006YA, SK514480007YA,SK514480008YA,SK514480009YA,SK514480010YA,SK514480011YA,SK514480012YA, SK514480013YA,SK514480014YA,SK514480015YA,SK514480016YA,SK514480017YA,SK514480018YA,SK514480019YA,SK514480020YA,SK514480021YA,SK514480022YA,SK514480023YA,SK514480024YA, SK514480025YA,SK514480026YA,SK514480027YA,SK514480028YA,SK514480029YA,SK514490018YA, SK514490019YA,SK514500001YA,SK514500002YA,SK514500003YA,SK514500004YA,SK514500005YA, SK514500006YA,SK514500007YA,SK514500008YA,SK514500009YA,SK514500011YA,SK515030069YA, SK515040001YA,SK515040002YA,SK515040003YA,SK515060004YA,SK515070014YA,SK515070015YA, SK515070016YA,SK515090008YA,SK515090009YA,SK515090010YA,SK515090011YA,SK515090012YA, SK515090013YA,SK515090014YA,SK515090015YA,SK515090016YA,SK515100035YA,SK515100036YA, SK515100037YA,SK515100038YA,SK515100039YA,SK515100040YA,SK515100041YA,SK515100042YA, SK515100043YA,SK515100044YA,SK515100045YA,SK515100046YA,SK515100047YA,SK515100048YA, SK515100049YA,SK515100050YA,SK515100054YA,SK515100056YA,SK515100057YA,SK515100058YA, SK515100059YA,SK515100060YA,SK515100061YA,SK515100063YA,SK515100064YA,SK515100066YA, SK515100067YA,SK515100068YA,SK515100069YA,SK515100070YA,SK515100071YA,SK515100072YA, SK515100073YA,SK515100074YA,SK515100075YA,SK515100077YA,SK515100078YA,SK515100079YA, SK515100080YA,SK515100082YA,SK515100083YA,SK515100085YA,SK515100087YA,SK515100088YA, SK515100094YA,SK515100096YA,SK515110001YA,SK515110004YA,SK515110005YA,SK515110006YA, SK515110008YA,SK515110009YA,SK515110010YA,SK515110011YA,SK515110013YA,SK515110019YA, SK515110021YA,SK515110022YA,SK515110023YA,SK515110024YA,SK515110025YA,SK515110026YA, SK515110027YA,SK515110029YA,SK515110031YA,SK515110032YA,SK515110036YA,SK515110037YA, SK515110038YA,SK515110039YA,SK515110040YA,SK515110042YA,SK515110044YA,SK515110046YA, SK515110048YA,SK515110049YA,SK515110054YA,SK515110055YA,SK515110056YA,SK515110057YA, SK515110058YA,SK515110061YA,SK515110067YA,SK515110068YA,SK515110075YA,SK515110076YA, SK515110077YA,SK515110078YA,SK515110079YA,SK515110080YA,SK515110081YA,SK515110082YA, SK515110083YA,SK515110084YA,SK515110085YA,SK515110086YA,SK515110087YA,SK515110088YA, SK515110089YA,SK515110090YA,SK515110091YA,SK515110092YA,SK515110116YA,SK515110117YA, SK515110118YA,SK515110119YA,SK515110120YA,SK515110122YA,SK515110123YA,SK515110124YA, SK515110125YA,SK515110126YA,SK515110127YA,SK515110128YA,SK515110129YA,SK515110130YA, SK515110131YA,SK515110132YA,SK515110133YA,SK515110134YA,SK515110135YA,SK515110136YA, SK515110138YA,SK515110139YA,SK515110140YA,SK515110141YA,SK515110142YA,SK515110143YA, SK515110144YA,SK515110145YA,SK515110146YA,SK515110147YA,SK515110148YA,SK515110149YA, SK515110150YA,SK515110151YA,SK515110152YA,SK515110153YA,SK515110154YA,SK515110155YA, SK515110156YA,SK515110157YA,SK515110158YA,SK515110159YA,SK515110160YA,SK515110161YA, SK515110162YA,SK515110163YA,SK515110164YA,SK515110165YA,SK515110166YA,SK515110168YA, SK515110169YA,SK515110170YA,SK515110171YA,SK515110172YA,SK515110175YA,SK515110176YA, SK515110177YA,SK515110178YA,SK515110181YA,SK515110197YA,SK515110198YA,SK515110199YA, SK515110200YA,SK515110202YA,SK515110203YA,SK515110204YA,SK515110205YA,SK515110206YA, SK515110207YA,SK515110208YA,SK515110214YA,SK515110219YA,SK515110222YA,SK515120010YA, SK515120018YA,SK515120028YA,SK515120030YA,SK515120031YA,SK515120033YA,SK515120034YA, SK515120035YA,SK515130004YA,SK515130005YA,SK515130006YA,SK515170019YA,SK515170022YA, SK514040029YA,SK514100003YA,SK514100005YA,SK514100006YA,SK514100007YA,SK514100010YA, SK514100011YA,SK514100013YA,SK514100014YA,SK514100015YA,SK514100030YA,SK514100031YA, SK514130002YA,SK514130004YA,SK514130005YA,SK514130010YA,SK514130016YA,SK514480032YA, SK514480033YA,SK514480034YA

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0264-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0264-2016Class IITerminatedVoluntary: Firm initiated