Z-0295-2016Class II

Fresenius 132 Gallon Dry Acid Dissolution Unit, P/N 160147. For Use in the Preparation of acid concentrate...

Fresenius Medical Care Renal Therapies Group, LLC / initiated 2015-10-23 / Terminated / root cause: design

Fresenius Medical Care Renal Therapies Group, LLC began a recall of Fresenius 132 Gallon Dry Acid Dissolution Unit, P/N 160147. For Use in the Preparation of acid concentrate for hemodialysis on . FDA classified it as Class II and gives the reason as "Dissolution units and replacement motors units do not provide full thermal protection for both the Main and Starter coils resulting in the generation of smoke and/or potentially lead to fire". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0295-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Fresenius Medical Care Renal Therapies Group, LLC
Product
Fresenius 132 Gallon Dry Acid Dissolution Unit, P/N 160147. For Use in the Preparation of acid concentrate for hemodialysis.
Product code
KPO Dialysate Concentrate For Hemodialysis (Liquid Or Powder)
Reason for recall
Dissolution units and replacement motors units do not provide full thermal protection for both the Main and Starter coils resulting in the generation of smoke and/or potentially lead to fire.
Root cause tag
design
FDA root cause
Device Design
Action
Fresenius issued an Urgent - Medical Device Correction notification, dated October 21, 2015, on 10/27/15. The letter identified the affected device and the reason for the recall. It also contained instructions (on how to perform the required wiring modification) along with a Reply Form. Customers are to return the form after completing the modification. Inquiries regarding the correction should be directed to Fresenius Technical Services at 800-227-2572, referencing the Field Action number FA-2015-01-W.
Quantity in commerce
348 units
Distribution
Nationwide Distribution.
Product codes and lots
All 132 Gallon Granuflo Dissolution Units and 132 Gallon Dry Acid Dissolution Units with the H218 Motor and Modified H218 Mixer Motors.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0295-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0294-2016Class IITerminatedVoluntary: Firm initiated
Z-0295-2016Class IITerminatedVoluntary: Firm initiated
Z-0296-2016Class IITerminatedVoluntary: Firm initiated