Fresenius 132 Gallon Dry Acid Dissolution Unit, P/N 160147. For Use in the Preparation of acid concentrate...
Fresenius Medical Care Renal Therapies Group, LLC began a recall of Fresenius 132 Gallon Dry Acid Dissolution Unit, P/N 160147. For Use in the Preparation of acid concentrate for hemodialysis on . FDA classified it as Class II and gives the reason as "Dissolution units and replacement motors units do not provide full thermal protection for both the Main and Starter coils resulting in the generation of smoke and/or potentially lead to fire". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0295-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Fresenius Medical Care Renal Therapies Group, LLC
- Product
- Fresenius 132 Gallon Dry Acid Dissolution Unit, P/N 160147. For Use in the Preparation of acid concentrate for hemodialysis.
- Product code
- KPO Dialysate Concentrate For Hemodialysis (Liquid Or Powder)
- Reason for recall
- Dissolution units and replacement motors units do not provide full thermal protection for both the Main and Starter coils resulting in the generation of smoke and/or potentially lead to fire.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Fresenius issued an Urgent - Medical Device Correction notification, dated October 21, 2015, on 10/27/15. The letter identified the affected device and the reason for the recall. It also contained instructions (on how to perform the required wiring modification) along with a Reply Form. Customers are to return the form after completing the modification. Inquiries regarding the correction should be directed to Fresenius Technical Services at 800-227-2572, referencing the Field Action number FA-2015-01-W.
- Quantity in commerce
- 348 units
- Distribution
- Nationwide Distribution.
- Product codes and lots
- All 132 Gallon Granuflo Dissolution Units and 132 Gallon Dry Acid Dissolution Units with the H218 Motor and Modified H218 Mixer Motors.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0294-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0295-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0296-2016 | Class II | Terminated | Voluntary: Firm initiated |