Z-0353-2016Class II
Philips Healthcare Xper cardio systems R7.6 Model : 722133
Philips North America LLC / initiated 2015-02-03 / Terminated / root cause: manufacturing process
Philips Medical Systems, Inc. began a recall of Philips Healthcare Xper cardio systems R7.6 Model : 722133 Product Usage: The Allura Xper series is intended for use in cardiovascular and vascular X-ray... on . FDA classified it as Class II and gives the reason as "Monitor Ceiling Suspension system may fall". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0353-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Philips Healthcare Xper cardio systems R7.6 Model : 722133 Product Usage: The Allura Xper series is intended for use in cardiovascular and vascular X-ray imaging applications including diagnostics, interventional procedures (such as PTCA, stent placement, embolization and thrombolysis and atherectomics), pacemaker Implantations and electrophysiology (EP). Use of the Allura Xper series system in the OR environment is restricted; the system should only be used for endo-vascular procedures. The Philips Integris Allura 9F is intended for diagnostic cardiovascular, vascular and interventional procedures
- Product code
- OWB Interventional Fluoroscopic X-Ray System
- Reason for recall
- Monitor Ceiling Suspension system may fall
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Philips sent an Urgent Field Safety Notice letter dated January 8, 2015 to affected customers. The letter identified the affected product, problem and actions to be taken. The actuator of all affected systems will be replaced free of charge. All affected customers will be contacted in order to schedule the implementation of this action. For questions contact your local Philips representatice: Technical Support Line: 1-800-722-9377.
- Quantity in commerce
- 14 units
- Distribution
- Worldwide Distribution - US Nationwide and the countries of: Canada,Australia, Austria, Brazil, Chile, China, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, South Africa, Spain, Switzerland and United Kingdom.
- Product codes and lots
- S/N: 118, 206, 207, 208, 211, 214, 223, 226, 234, 244, 248, 261
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0346-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0347-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0348-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0349-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0350-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0351-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0352-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0353-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0354-2016 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.