Z-0222-2016Class II

Stryker UniVise Spinous Process Fixation Plate System Inserter, Stryker Spine, Non-Sterile. Reference Number...

Howmedica Osteonics Corp / initiated 2015-09-08 / Terminated / root cause: manufacturing process

Stryker Spine began a recall of Stryker UniVise Spinous Process Fixation Plate System Inserter, Stryker Spine, Non-Sterile. Reference Number 48590100. Manual Surgical Instrument on . FDA classified it as Class II and gives the reason as "The Inserter Inner Shaft would not fit through the Inserter. The inner diameter of the cannulated shaft of the inserter was undersized". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0222-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Howmedica Osteonics Corp
Product
Stryker UniVise Spinous Process Fixation Plate System Inserter, Stryker Spine, Non-Sterile. Reference Number 48590100. Manual Surgical Instrument.
Product code
LXH Orthopedic Manual Surgical Instrument
Reason for recall
The Inserter Inner Shaft would not fit through the Inserter. The inner diameter of the cannulated shaft of the inserter was undersized.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Stryker Spine issued an "Urgent Product Removal Notifications /Customer Response Forms dated September 11, 2015. to all affected customers via Fed Ex priority overnight. The notification instructed customers on how to identify affected product; the related issue with the affected product; potential hazards; risk mitigations; and actions needed to be taken. Customers were asked to examine their inventory and related hospital location inventories; and reconcile any UniVise Inserters with Catalog numbers and Lot numbers (referenced in the notification). Customers were asked to fax a copy of the Customer Response Form to Regulatory Compliance (at 855-632-9049) or email it to [email protected] within 5 days of receipt of the notification. For further questions please call (201) 760-8346.
Quantity in commerce
41 units
Distribution
Nationwide Distribution.
Product codes and lots
Reference Number 48590100 Lots B40155, B34446, B34447 and B41171.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0222-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0222-2016Class IITerminatedVoluntary: Firm initiated