Z-0250-2016Class II

NovoCut Suture Manager; Ceterix catalog number CTX-001. General and Plastic Surgery: Intended for use in...

Ceterix Orthopedics, Inc. / initiated 2015-10-13 / Terminated / root cause: design

Ceterix Orthopedics, Inc. began a recall of NovoCut Suture Manager; Ceterix catalog number CTX-001. General and Plastic Surgery: Intended for use in arthroscopic surgery on . FDA classified it as Class II and gives the reason as "The NovoCut Suture Manager device may cut the suture during knot tensioning and this may cause the suture knot to unravel". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0250-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ceterix Orthopedics, Inc.
Product
NovoCut Suture Manager; Ceterix catalog number CTX-001. General and Plastic Surgery: Intended for use in arthroscopic surgery.
Product code
NBH Accessories, Arthroscopic
Reason for recall
The NovoCut Suture Manager device may cut the suture during knot tensioning and this may cause the suture knot to unravel.
Root cause tag
design
FDA root cause
Device Design
Action
Ceterix Orthopaedics sent an Urgent Field Safety notification letter dated October 13, 2015, to all affected sites. The product was identified, reason for recall explained and letter contains list of product lot numbers affected. Customers were instructed to identify and quarantine the affected product, contact Ceterix Orthopaedics to arrange a no cost return, and complete and return the attached acknowledgement form. Customers with questions were instructed to call 1-888-305-0615 or email: [email protected]. For questions regarding this recall call 650-396-3478.
Quantity in commerce
1364 units
Distribution
Nationwide Distribution
Product codes and lots
Ceterix catalog number CTX-001; Lot numbers: M150016 M150056 M150078 M150105 M150136 M150182 M150185 M150202 M150207 M150225 M150239.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0250-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0250-2016Class IITerminatedVoluntary: Firm initiated