510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0443-2016Class II

Oxoid Gentamicin Antimicrobial Susceptibility Testing Disc (CN10), REF# CT0024B packaged in packs of 5...

Remel Inc / initiated 2015-11-06 / Terminated / root cause: manufacturing process

Remel Inc began a recall of Oxoid Gentamicin Antimicrobial Susceptibility Testing Disc (CN10), REF# CT0024B packaged in packs of 5 cartridges - 50 disc/cartridge. Used in the... on . FDA classified it as Class II and gives the reason as "Antibiotic concentration may be insufficient which may result in incorrect test results". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0443-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Remel Inc
Product
Oxoid Gentamicin Antimicrobial Susceptibility Testing Disc (CN10), REF# CT0024B packaged in packs of 5 cartridges - 50 disc/cartridge. Used in the semi-quantitative agar diffusion test method for in vitro susceptibility testing to assist in the determination of a suitable antibiotic agent for treatment purposes.
Product code
JTN Susceptibility Test Discs, Antimicrobial
Reason for recall
Antibiotic concentration may be insufficient which may result in incorrect test results.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
ThermoFisher Scientific sent an Urgent Medical Device Recall letter dated November 6, 2015, to all affected customers via first class mail. The recall notice identified the problem with the device and asked customers to inspect their stocks for the affected lots. The notice instructed customers to destroy any remaining inventory from the affected lots, contact the recalling firm for replacement, and complete the accompanying Acknowledgement form. Customers with questions were instructed to contact the Technical Services Departmentn at 800-255-6730 (US) or 913-888-0939 (International). For questions regarding this recall call 913-895-4077.
Quantity in commerce
58/5/50 discs/catridges/packs
Distribution
Nationwide Distribution including IL, GA, FL, NC, CA, AL, MA, IN, MO, NY, TX, VA, KS, and IA. Product was also distributed to Puerto Rico.
Product codes and lots
Lot 1686823, expiry 30 June 2018

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0443-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0443-2016Class IITerminatedVoluntary: Firm initiated