Z-0321-2016Class II
Tool Legend 16cm 2.4 mm wire pass for Midas Rex Legend Dissecting Tool. Used by surgeons to drill holes
Medtronic Powered Surgical Solutions / initiated 2015-10-21 / Terminated / root cause: manufacturing process
Medtronic Sofamor Danek Usa, Inc - Dallas Distribution began a recall of Tool Legend 16cm 2.4 mm wire pass for Midas Rex Legend Dissecting Tool. Used by surgeons to drill holes on . FDA classified it as Class II and gives the reason as "The wire pass hole is not present on the tools". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0321-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic Powered Surgical Solutions
- Product
- Tool Legend 16cm 2.4 mm wire pass for Midas Rex Legend Dissecting Tool. Used by surgeons to drill holes.
- Product code
- HBB Motor, Drill, Pneumatic
- Reason for recall
- The wire pass hole is not present on the tools.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- A recall letter was mailed to consignees on October 21, 2015.
- Quantity in commerce
- 9 pieces
- Distribution
- Distributed in the states of California, New York, and Texas and the country of Belgium.
- Product codes and lots
- Catalog Number: 16TA24, Lot Number: 0006936810