Z-0563-2016Class II

Ceiling Mounted Surgical Light System TruLight 3000/5000 and Ceiling Mounted Surgical Light System iLED3/iLED5

Hill-Rom Inc. / initiated 2015-11-02 / Completed

Trumpf Medical Systems, Inc. began a recall of Ceiling Mounted Surgical Light System TruLight 3000/5000 and Ceiling Mounted Surgical Light System iLED3/iLED5 on . FDA classified it as Class II and gives the reason as "The firm has received 15 complaints over ten years related to the connection of the spring arm to the top horizontal arm of the lighting system that while positioning the light head enabled the spring arm and light head...". As of , the recall status on record is Completed.

Recall record

Recall number
Z-0563-2016
Classification
Class II
Status
Completed
Initiated
Posted
Recalling firm
Hill-Rom Inc.
Product
Ceiling Mounted Surgical Light System TruLight 3000/5000 and Ceiling Mounted Surgical Light System iLED3/iLED5
Product code
FSY Light, Surgical, Ceiling Mounted
Reason for recall
The firm has received 15 complaints over ten years related to the connection of the spring arm to the top horizontal arm of the lighting system that while positioning the light head enabled the spring arm and light head to fall.
FDA root cause
Other
Action
TRUMPF sent an " Urgent Field Safety Information" letter dated December 8, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Actions to be taken by the customer: 1. Perform an inspection of the lighting systems (see Appendix 1 for instructions). If any of the symptoms listed below are observed, immediately remove the light from operation and contact your Trumpf Medical Service Representative or designee. Signs that indicate a possible fall of the spring arm: The movement of the light and spring arm is difficult for the user, The lights have stopped being operational, and/or A gap is noticed between the spring arm and the central axis (see the attached Appendix 1 for an example). 2. Follow the User Manual: No unauthorized servicing of the spring arm is permitted. Snap rings are one use only and replacements must be purchased from your authorized Trumpf Medical partner. If you have any questions concerning this request or the procedure as outlined, please contact TRUMPF Medical Systems, Inc. at 888-474-9359
Quantity in commerce
20,713
Distribution
USA (nationwide) Distribution to the states of : AL, AK, AZ, CA, CO, CT, DE, FL, GA, ID, IL, IN, IA, KY, ME, MA, MN, MO, NE, NV, NJ, NM, NY, NC, ND, OH, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI and WY., and Internationally to Canada.
Product codes and lots
4028110 - iLED Single; 4028210 - iLED duo; 4028310 - iLED trio/quad; 4038110 - TruLight single; 4038210 - TruLight duo; 4038310 - TruLight trio/quad; 1565068 - iLED single (TUSM); 1565160 - iLED duo (TUSM); 1565161 - iLED trio/quad (TUSM); 1574759 - TruLight single (TUSM); 1574850 - TruLight duo (TUSM); 1574851 - TruLight trio/quad (TUSM)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0563-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0563-2016Class IICompletedVoluntary: Firm initiated